background image

11

CAUTION

Storage of the product in the closed state

Injury  due  to  faulty  control  or  malfunction  of  the  product  as  a  result  of  damage  to  the  sensors
and mechanism.

Store the product only in the neutral position or in the open state.

1) Push and hold the charging receptacle button until a confirma­
tion sound is heard (at least 1 second).

The prosthesis and prosthetic components are switched on.

Repeat this process to switch the prosthesis and prosthetic compo­
nents off.

INFORMATION

The Axon-Bus prosthetic system cannot be used as long as the charging plug is connected to
the charging receptacle. The Axon-Bus prosthetic system is deactivated for the duration of char­
ging.

7.3 Opening the Axon-Bus gripping component in an emergency

This safety function allows the Axon-Bus gripping component to be
opened regardless of the control signals present.
1)  With the Axon-Bus prosthetic system switched on, press the char­

ging receptacle button and hold for approximately three seconds
until the Axon-Bus gripping component begins to open. 

A pulsating beep sounds as it opens.

2)  Releasing  the  button  immediately  stops  the  process  of  opening

the Axon-Bus gripping component and turns the entire Axon-Bus
prosthetic system off.

8 Maintenance

Performing  regular  maintenance  (service  inspections)  every  24  months  is  recommended  to  pre­
vent injuries and maintain the quality of the product.
In general, all products are subject to compliance with the maintenance intervals during the war­
ranty period. This is the only way to maintain full warranty cover.
Additional services such as repairs may be provided in the course of maintenance. These addi­
tional  services  may  be  provided  free  of  charge  or  can  be  billable  according  to  an  advance  cost
estimate, depending on the extent and validity of the warranty.
The following components must always be sent in for maintenance and repairs:
The  product,  battery  charger  and  power  supply.  The  shipping  container  for  the  loaner  unit  you
receive must be reused for sending back the components requiring inspection.

9 Legal information

9.1 Liability

The manufacturer will only assume liability if the product is used in accordance with the descrip­
tions  and  instructions  provided  in  this  document.  The  manufacturer  will  not  assume  liability  for

Содержание Michelangelo Hand 8E500

Страница 1: ...Instructions for use qualified personnel 3 Michelangelo Hand 8E500...

Страница 2: ...1 2 2...

Страница 3: ...t into use in accordance with the information contained in the accompanying documents supplied 2 Product description 2 1 Function The Michelangelo Hand is a myoelectrically controlled gripping compone...

Страница 4: ...mately 75 in 4 locking positions extension is 45 with 3 locking positions Flexion and extension are performed passively A flexible and a rigid mode are available according to the application situation...

Страница 5: ...3 Indications Amputation level transradial and transhumeral For unilateral or bilateral amputation Dysmelia of the forearm or upper arm The user must be able to understand usage instructions and safe...

Страница 6: ...active implanted systems Interference with active implantable systems e g pacemaker defibrillator etc due to electro magnetic interference of the prosthetic system When operating the prosthetic system...

Страница 7: ...tware Injury due to unexpected product behaviour Participation in an Ottobock product training course is mandatory prior to using the product During this product training course you will receive a pas...

Страница 8: ...nents e g the terminal device CAUTION Changing Axon Bus gripping components when switched on Injury due to faulty control or malfunction of the Axon Bus prosthetic system Power down the Axon Bus prost...

Страница 9: ...magnetic data carriers Wiping of the data carrier Do not place the charging plug on credit cards diskettes audio or video cassettes 5 Scope of Delivery and Accessories 5 1 Scope of delivery 1 pc 8E500...

Страница 10: ...nformation see the instructions for use provided with the 13E500 AxonMaster and the 560X500 AxonSoft software 7 Use 7 1 Changing the Axon Bus gripping component Attaching the Axon Bus gripping compone...

Страница 11: ...n A pulsating beep sounds as it opens 2 Releasing the button immediately stops the process of opening the Axon Bus gripping component and turns the entire Axon Bus prosthetic system off 8 Maintenance...

Страница 12: ...any interference received including interference that may cause undesired operation This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to part 15...

Страница 13: ...Canada limits for the general population consult Safety Code 6 obtainable from Health Canada s website http www hc sc gc ca rpb Responsible party Otto Bock Healthcare Canada Ltd 5470 Harvester Road L...

Страница 14: ...N Gripping force Lateral Mode Approx 60 N Gripping force Neutral Mode Approx 15 N Load limits Maximum vertical load on the palm with locked wrist joint e g when holding a sphere 10 kg 22 1 lbs Maximu...

Страница 15: ...15...

Страница 16: ...663 339 US 8 690 963 European Patent EP 1962732 in AT DE FR GB IT NL SE TR EP 1962734 in AT DE FR GB IT NL SE TR EP 1962737 in AT DE FR GB IT NL SE TR EP 1962738 in DE FR GB TR EP 1971297 in AT DE FR...