Safety modes
–
Severe error
•
Loss of a sensor or the AXON tube adapter
•
These errors are displayed with
30 long beep signals and
alternating short and long vibration signals
•
Stance flexion resistance corresponds to the ‘Activity Mode’:
•
Mode A / B / B+:
•
Knee joint with very high flexion resistance
→
Allows safe standing and to flex the knee joint during
sitting down
•
Mode C:
•
Adjusted stance flexion resistance
•
Stance release is not possible
•
Try to fix the error by connect and disconnect the charger (run
the self-test)
See your CPO
immediately!
If the error is present after several self-tests, the usage of the knee joint is
not permitted anymore and it must be sent back to Ottobock!