CAUTION
Penetration of dirt and humidity into the product
Risk of injury due to unexpected product behaviour or malfunction.
►
Ensure that neither solid particles nor liquids can penetrate into the product.
►
Do not expose the product, and especially the elbow joint, to dripping or splashing water.
►
Wear the product, and especially the elbow joint, under suitably resistant clothing in the rain.
CAUTION
Changing prosthetic components when switched on
Injury due to faulty control or malfunction of the prosthetic components.
►
Before changing prosthetic components (e.g. terminal device) switch the product off.
CAUTION
Signs of wear and tear on the product components
Injury due to faulty control or malfunction of the product
►
In the interest of patient safety and in order to maintain operating reliability and protect the warranty, the spe
cified service intervals must be observed (see the section "Maintenance").
CAUTION
Injury due to incorrect adjustment of the product.
Loosening of the components.
►
In case of a product with a quick-disconnect wrist, position the terminal device before use so that a slight turn
cannot disconnect the terminal device from the prosthesis.
CAUTION
Incorrect donning and doffing of the product
Injuries
►
The product should always be flexed for donning and doffing as well as storage.
►
Switch off the product before donning or doffing.
CAUTION
Unintentional unlocking of the terminal device
Risk of injury due to releasing the terminal device from the forearm (e.g. while carrying objects).
►
When using a quick-disconnect wrist unit, position the terminal device so that slight twisting does not release
the terminal device from the forearm.
CAUTION
Charging the product with damaged power supply unit/charger/charger cable
Injury due to unexpected behaviour of the product caused by insufficient charging.
►
Check the power supply unit, charger and charger cable for damage before use.
►
Replace any damaged power supply unit, charger or charger cable.
CAUTION
Risk of pinching in the joint flexion area
Injuries due to pinching of body parts.
►
Ensure that fingers and other body parts are not in this area when bending the joint.
NOTICE
Coating, gluing or painting the prosthesis
Damage or fracture due to chemical processes.
►
The prosthesis must not be coated, glued or painted.
8
Safety
DynamicArm 12K100N
Содержание DynamicArm 12K100N
Страница 1: ...DynamicArm 12K100N Instructions for use qualified personnel 3...
Страница 30: ...30 DynamicArm 12K100N...