9 Maintenance
CAUTION
Failure to follow the maintenance instructions
Risk of injuries due to changes in or loss of functionality and dam
age to the product
►
Observe the following maintenance instructions.
►
Arrange regular maintenance intervals with the patient depending
on the level of use.
►
Following an individual period for the patient to get accustomed
to the prosthesis, check the settings of the prosthetic joint and
adapt them to the patient's requirements again as needed.
►
The prosthetic components should be inspected after the first 30
days of use.
►
Inspect the entire prosthesis for wear during normal consulta
tions.
►
Conduct annual safety inspections.
►
As part of the safety inspections, inspect the prosthetic joint for
wear and proper functionality. Special attention should be paid to
movement resistance, bearings and abnormal noises. Full flexion
and extension must be ensured at all times. Readjust as required.
Lubrication
►
14
on page 8: Do not lubricate or grease the prosthetic joint.
Replacing components
CAUTION
Improper assembly of the screw connections
Risk of injury due to breakage or loosening of the screw connec
tions
►
Clean the threads before every installation.
►
Apply the specified torque values.
►
Follow the instructions regarding the length of the screws and
about how to secure the screws.
15
►
When
necessary,
remove
a
worn
brake
bushing
(4B74=20.4X24.5-K) as shown in the illustration
15
on page 9.
►
To reassemble, perform these steps in reverse order.
►
In doing so, note the following points:
→
Ensure the correct assembly of the bushing safety device.
→
Clean the knee axis with isopropyl alcohol ( 634A58)
before assembly.
10 Disposal
In some jurisdictions it is not permissible to dispose of the product
with unsorted household waste. Improper disposal can be harmful to
health and the environment. Observe the information provided by the
responsible authorities in your country regarding return, collection
and disposal procedures.
11 Legal information
11.1 Liability
The manufacturer will only assume liability if the product is used in
accordance with the descriptions and instructions provided in this
document. The manufacturer will not assume liability for damage
caused by disregarding the information in this document, particularly
due to improper use or unauthorised modification of the product.
11.2 CE conformity
The product meets the requirements of Regulation (EU) 2017/745
on
medical devices. The CE declaration of conformity can be down
loaded from the manufacturer's website.
1 Информация по данному документу
Русский
ИНФОРМАЦИЯ
Дата последней актуализации: 2021-01-22
Содержание 3R15
Страница 1: ...Instructions for use 10 15 3R15 3R49...
Страница 9: ...6 mm 634A58 Isopropyl alcohol 4B74 20 4X24 5 K 25 Nm 0 mm 9 Maintenance 15 9...
Страница 16: ...1 1 1 2 1 1 1 J 2 3R15 3R49 16 3R15 3R49 3 3 1 3 2 3R15 1 2 100 kg 1...
Страница 17: ...2 3 3 17 3 4 10 C 60 C 0 90 100 250 1100 3 5...
Страница 18: ...3 5 4 18 5 1 2 6 2 2 PSH DSH 7 7 1...
Страница 19: ...7 2 3 3 4 3 5 4 7 3 19 6 4 Spina iliaca anterior superior 6 4 7 4 7 8 9 5 6 9 6 45...
Страница 20: ...10 6 45 90 11 7 20 12 7 20 7 5 13 8...
Страница 21: ...7 5 1 3R6 3S106 8 8 1 21 8 2 1 2 3...
Страница 22: ...9 30 14 8 22 4B74 20 4X24 5 K 15 9 634A58 10 11 11 1...
Страница 23: ...11 2 2017 745 CE 23...