If the extension position is not reached with every step and at any
walking speed, reduce the braking effect and/or increase the ini
tial tension of the extension assist.
►
Check the adjustment of the prosthesis settings during normal
consultations and the annual safety inspection.
Advise the patient to check the prosthesis if changes in function
occur.
7.5 Finishing the prosthesis
Put on protective cover
►
When the prosthesis is finished, put on the protective cover as
shown in the illustration
13
on page 8.
: Alternative positioning
: Recommended positioning
►
CAUTION!
To avoid product damage and the risk of falling, finish
the prosthesis by replacing set screws that are too short or too
long, and by tightening all screw connections. In doing so, note
the instructions for use for all prosthesis components regarding
torque values and thread lock.
7.5.1 Attaching the cosmetic cover
CAUTION
Use of talcum
Risk of injury, damage to the product due to lack of lubrication
►
Do not use talcum on the product or other prosthesis compon
ents.
Recommended combinations
Foam cover
3R6, 3S106
►
After affixing the cosmetic elements, verify proper functioning of
the prosthesis.
►
To reduce friction and to eliminate noise, apply silicone spray dir
ectly onto the contact surfaces of the cosmetic foam cover.
14
8 Use
8.1 Information for use
NOTICE
Mechanical overload
Impaired functionality due to mechanical damage
►
Check the product for damage prior to each use.
►
Do not use the product if its functionality has been impaired.
►
Take any necessary measures (e.g. repair, replacement, inspec
tion by the manufacturer's customer service, etc.).
8.2 Cleaning
CAUTION
Use of unsuitable cleaning agents or disinfectants
Impairment of functionality and damage due to incorrect cleaning
agents or disinfectants
►
Clean the product only according to the instructions given in this
section.
►
For the prosthesis, observe the cleaning instructions for all pros
thetic components.
INFORMATION
►
Clean the product if it gets dirty.
►
For disinfection, only use disinfectants that do not affect the
materials of the product. For further information, please consult
the manufacturer.
►
Observe the cleaning instructions for all prosthetic components.
1) Clean the product with a damp, soft cloth.
2) Dry the product with a soft cloth.
3) Allow to air dry in order to remove residual moisture.
Содержание 3R15
Страница 1: ...Instructions for use 10 15 3R15 3R49...
Страница 9: ...6 mm 634A58 Isopropyl alcohol 4B74 20 4X24 5 K 25 Nm 0 mm 9 Maintenance 15 9...
Страница 16: ...1 1 1 2 1 1 1 J 2 3R15 3R49 16 3R15 3R49 3 3 1 3 2 3R15 1 2 100 kg 1...
Страница 17: ...2 3 3 17 3 4 10 C 60 C 0 90 100 250 1100 3 5...
Страница 18: ...3 5 4 18 5 1 2 6 2 2 PSH DSH 7 7 1...
Страница 19: ...7 2 3 3 4 3 5 4 7 3 19 6 4 Spina iliaca anterior superior 6 4 7 4 7 8 9 5 6 9 6 45...
Страница 20: ...10 6 45 90 11 7 20 12 7 20 7 5 13 8...
Страница 21: ...7 5 1 3R6 3S106 8 8 1 21 8 2 1 2 3...
Страница 22: ...9 30 14 8 22 4B74 20 4X24 5 K 15 9 634A58 10 11 11 1...
Страница 23: ...11 2 2017 745 CE 23...