Option
Description
Files
Used to view or delete files.
Setup
Used to set up filter parameters and the Maximum depth feature.
3.
Select Setup > Maximum depth. Select Manual or Automatic.
Option
Description
Manual
In this mode, the instrument prompts the user to manually enter the maximum depth of each
vertical. This value is normally obtained from a wading rod.
Automatic
In this mode (available only on sensors with the depth option), the instrument uses the pressure
transducer to measure the maximum depth at each vertical.
If Automatic is selected, enter the distance from the bottom of the channel to the bottom of the
sensor mount (Offset). The Offset distance value can be different based on the used wading rod
and the properties of the bottom. Refer to
Figure 12
. When rods with a base plate and/or tip are
used, this extra distance can be larger. This depends on whether the subsurface is firm or yields.
The extra distance is important as a parameter for the correct determination of the water depth.
Figure 12 Offset distance (D)
4.
If necessary, change or update the filter parameters in Profiler Setup.
5.
In the Profiler menu, select Stream.
6.
Enter the operator name. Make operator names alpha-numeric with a maximum of 23 characters.
Push
OK
to save the operator name.
7.
Enter the name of the new stream profile. Make profile names alpha-numeric with a maximum of
23 characters. Push
OK
to save the profile name or select Clear to delete all current stream data.
8.
In the Station menu, select Edge/Obstruction. Select one of the options.
Option
Description
Left
Select this option if the station is at the left edge of the water or an obstruction (i.e, sandbar,
pylon or large boulder).
22
English