Figure 9 Real time screen for sensor with depth
2. Profiler—
Select this option to do stream or conduit velocity measurements. The meter shows
prompts when user input is necessary. A percentage of the remaining memory is given in 1%
resolution. Refer to the
Appendix
on page 31 for more information about profiles.
Option
Description
Stream
This option is used to set up a profile for a stream or flowing channel. Do velocity measurements to
calculate total discharge based on ISO 748 or USGS standards for Mid-section or Mean-section
methods.
Conduit
This option is used to set up a profile for a conduit.
Files
This option is used to view or delete stored files. Files can be deleted all at once or individually.
Setup
This option is used to set up or change the settings for filter parameters and the Maximum Depth
feature.
3. Set up
—Select this option to change general system settings and preferences.
Option
Description
Velocity
calibration
Calibrates the sensor. Adds a field offset to the factory calibration. Refer to the
Appendix
on page 31 for more information.