13.2
Notes on EMC (Electromagnetic Compatibility)
•
MADSEN Zodiac is part of a medical electrical system and is thus subject to special safety precautions. For this reason,
the installation and operating instructions provided in this document should be followed closely.
•
Portable and mobile high-frequency communication devices, such as mobile phones, may interfere with the func-
tioning of MADSEN Zodiac.
IEC 60601-1-2:2014 and EN 60601-1-2:2015
Guidance and manufacturer's declaration - electromagnetic emissions for all equipment and systems
MADSEN Zodiac is intended for use in the electromagnetic environment specified below. The user of MADSEN Zodiac should ensure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
MADSEN Zodiac uses RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
MADSEN Zodiac is suitable for use in all environments, including domestic environments and
those directly connected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions IEC
61000-3-2
Not applicable
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Not applicable
Guidance and manufacturer's declaration - electromagnetic immunity for all equipment and systems
MADSEN Zodiac is intended for use in the electromagnetic environment specified below. The user of MADSEN Zodiac should ensure that it is used in such an
environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment - guidance
Otometrics - MADSEN Zodiac
User Guide
32