Press the soft-button below the next column to select this column, and scroll to any setting you wish to view or
change.
How to change test settings
1. Start by selecting a protocol.
2. Select the center column.
3. Scroll to the setting you wish to view or change.
4. Select the right-hand column to change the setting.
5. Scroll to the value you wish to use and press the
Select
soft-button.
6. Repeat steps 2 to 5 to change other settings, if needed.
Using test settings temporarily
•
If you only wish to use the changed protocol settings temporarily, select the left-hand column, scroll to
Apply Tem-
porarily
and press the
OK
soft-button. This will exit the menu system.
Saving test settings in the protocol
Before you can save test settings in the protocol, you must unlock the protocol: Select
Device Settings..
>
Protocol Set-
tings
, and set to
Allowed
. If you switch off Zodiac after changing and saving the test settings, the protocol will be locked
again.
•
If you wish to save the changed protocol settings permanently, select the left-hand column, scroll to
Save to [Protocol
name]
, and press the
Save
soft-button.
Canceling changed test settings
•
If you wish to cancel the settings you have changed, select the protocol and press the
Reload
soft-button.
8
Testing with Zodiac Quick Check - Stand-alone
Note
•
The safety intensity level cannot be exceeded in screening mode.
MADSEN Zodiac Quick Check performs automatic 226 Hz tym-
panometry which can be combined with Ipsi Reflex testing.
All tests are performed in one single sequence.
MADSEN Zodiac is used in combination with the hand-held
Quick Check probe fitted with an eartip.
The measurements are performed automatically when the
probe is pressed gently against the patient's ear canal and her-
metic seal is achieved.
8.1
Creating a new session
When you wish to test a new patient, create a new session. This will clear the device memory of old data.
Otometrics - MADSEN Zodiac
User Guide
14