Running operation
Before beginning the surgical operation, switch on the unit and wait for the thermal cleaning process to complete.
After the device had been idle for a long period of time (weekend, during holidays, etc.), the thermal cleaning process
must be initiated and repeated.
We recommend that the device be switched off and on every 2 hours and not longer than 4 hours to start the thermal
cleaning mode.
Immediately replace and autoclave contact parts such as the jet connection each time it is used for a different patient (for
prevention of retrograde contamination).
Before every use, spray water and check the temperature.
Service
Ensure that the suspended particle filter is hygienically changed (see commissioning) (dispose of filter and carefully clean
the filter glass). Ensure that no contact is made with the contaminated parts.
Carry out thermal cleaning after each service.
We recommend that the germ count in the water be regularly determined at the input and output of the unit.
3
Installation and start-up
3.1
Standard delivery
Quantity
1
Aquamatic™ II basic unit
1
Irrigation handle
1
Double-barreled hose
3
Jet connection short, straight (80 mm)
1
Splash protection
1
Rubber bush for splash protection
2
Three-hole gasket
1
Hose tips for jet connection (30 Pcs)
1
G3/4i-G1/4a-screwing (water-tap G3/4 on filter G1/4)
1
Cap nut G3/4
1
Water filter complete
2
Gasket (f.G1/4a) 13x18x1
1
G1/4a-G3/4a-reducer (filter G1/4 on water hose G3/4)
1
Water supply hose G3/4i, L = 3 m
1
Waste water hose G1/4i, L = 3 m