164
Quality checks
CHAPTER SEVENTEEN
Quality checks
Details
The full test result can be viewed by selecting
Details
. The details screen
shows the levels recorded from the probe loud speaker at each frequency
tested. The NEW column shows the levels just recorded and the OLD
results are the levels that are stored in the probe.
Results are given for each frequency tested:
Pass
- Tick/Check mark ( ) The
NEW
and
OLD (
Stored) data for each of
the two channels are within ± 3dB and are within the absolute limits.
Query
- Question mark ( ) Values differ by more than ± 3dB. The NEW
and OLD levels are highlighted.
Fail
- Cross ( ) Values are outside the absolute range. The NEW level
only is highlighted.
The 1, 2 and 4kHz values may not be stored in the probe if a new probe is
being used with the system. To save new data, run a Probe Test, record the
values for each frequency and repeat by selecting
Retest.
Check that the
values from two sequential tests are within ± 0.5dB before selecting
Save
.
Save
is only available to admin users. It is not available if the test was
noisy or if the levels were outside the absolute range.
On selecting
Save
, the screen title
Overwrite Stored?
will be shown
highlighted. Select
Yes
to save the new data or
No
to keep the current
stored values which may be blank for a newly registered probe. Before
entering Probe Test the user will be prompted to register the probe with the
Otoport.
Select
Back
to exit the
Probe Test
screen and return to the
Configuration
screen.