CHAPTER FIFTEEN
Configuration
122
Configuration
15.2.4
TEOAE
The
TE Quick
setting is used to switch the TE mode
on
(default) and
off
.
When
on
, the mode will provide a Quickscreen click stimulus of
84dB
. If it
is set to
off
, this test mode will not be available.
Mode
Mode
sets the pass criteria for screening.
The
Wide Band
setting requires a signal to noise ratio (SNR) of 6dB over
the 841-4757Hz frequency range in order to meet the pass criteria.
The
Narrow
setting also requires a signal to noise ratio (SNR) of 6dB
over the 841-4757Hz frequency range in order to meet the pass criteria.
However this setting uses a filter of 1.6-3.2kHz and a stimulus response
window 4-10ms to reduce the impact of environmental noise on the
recording. This makes this setting particularly appropriate for screening in
noisy environment.
The
Any 2 Bands
,
Any 3 Bands
and
Any 4 Bands
settings require a 6dB
SNR and a minimum signal level of -5dBSPL in each of the stated number
of half octave frequency bands to meet the pass criteria.
In order to meet the pass criteria, all test modes require a minimum total
(wide band) signal level of 0dBSPL and the amount of data collected to
reach the minimum number of sweeps.