There are two ways to add a spectrum to a library. The first one, open your spectrum, 

click the 

Add to library

 button, then select the library where you want to store your spectrum. 

You can only add spectra to user libraries, so you should create one beforehand. A library 

cannot contain same spectra, so your attempt to add the same spectrum twice will result in 

an error message. 

The second way is to drag your spectrum from its folder to the library. You should open 

a user library in one list, and spectrum folder in another, then drag one of the spectra to the 

first list. 

 

Troubleshooting 

All troubles listed below are described with their names, manifestations, causes and 

remedies. If none helps, please contact your supplier. 

1.

 

Fault:

 no connection to the device.

 Manifestation:

 in case of a single error, the 

program shows a message "

Failed to connect to the device...

" When the program 

is restarted and there is still no connection, the message will not show up in 

order not to disturb the user if the latter does not use the device on purpose, 

however, the window bottom right corner will display a disconnected connector 

icon instead of a shaking green spot. The log will be supplemented with line 

ft801: r= -1

... 

Possible causes:

 (a) the device is off, (b) there is no connection 

through the USB cable or it is unreliable, (c) 

IROS

 device driver is missing, (d) 

another instance of 

ZaIR 3.5

 is running and occupying the device. 

Remedies:

 

First, make sure that the device is on and connected to the computer with a 

cable. Then restart the program. If it does not help, unplug and plug the USB 

cable back in. Next, turn off the device, wait for a minute, turn it on again, and 

restart the program. If it does not help, restart the computer. 

2.

 

Fault:

  scan failure. 

Manifestation: 

the green spot at bottom right will be 

replaced with an icon corresponding to the error type, and the log will be 

supplemented with its code. Most often, this is a short scan (Code 5) or weak 

laser signal (Code 3). 

Possible causes: 

device case shaking, workplace 

vibrations etc.

 Remedies:

 do not subject your device to any mechanical impact, 

make sure that the table is stable. If faults occur regularly despite no apparent 

reason, especially increasing when working with high resolutions, the device 

might need adjusting by the manufacturer. 

3.

 

Fault:

 connection with the device lost. 

Manifestation:

 after a period of normal 

device operation, appears a message: "

Lost connection with the device...

Possible causes:

 (a) electromagnetic interference with the device-to-computer 

communication channel, (b) computer overloaded, the device fails to transfer 

data in time. 

Remedies: 

restart the program or unplug and plug in the USB port 

again. If lost connections keep occurring, you might need a shorter or ECM-

resistant cable, place it a maximum distance from the other devices. 

4.

 

Fault:

 no data transfer to the computer. 

Manifestation:

 the green light on the 

device is flashing, the interferometer carriage is moving, but the spot in the 

software window stands still with no error messages. If you click the spectrum-

recording button, the result will not appear in the window. 

Possible causes:

 

device driver data overrun or device fault. 

Remedies:

 unplug and plug the USB 

cable back in or, at worst, turn the device off for a while. 

 

 

Содержание 4434-801-59962935-08

Страница 1: ...IROS 05 IR Fourier Spectrometer Operating Manual FSVE 443430 005RE 2020 Ostec Instruments Moscow...

Страница 2: ...specialists who have read this manual are qualified to work on electrical devices up to 1 000 V and briefed on workplace safety rules The spectrometer involves operations with hazardous 220 V 50 Hz c...

Страница 3: ...n 6 Device software installation 6 First run 7 Spectrum recording 7 Spectrum reading and copying 9 Spectrum identification 12 Viewing spectra 15 Spectrum editing 18 Calibration 19 Creating and editing...

Страница 4: ......

Страница 5: ...434 801 59962935 08 Please see spectrometer operating conditions in Table 1 Spectrometer s electromagnetic interference resistance complies with GOST 30804 6 1 2013 and radiated interference with GOST...

Страница 6: ...ise ratio RMS in standard conditions in1 min at a resolution of 4 cm 1 in full bandwidth of 2000 2200 cm 1 40 000 Signal gain 1 2 4 8 Resolution 0 5 1 2 4 8 cm 1 Scanning speed spectra per minute For...

Страница 7: ...1 Spectrometer optical scheme The inlet optical system 2 converts the IR source 1 radiant energy into a convergent annular beam and makes it focus on the inlet port of the interferometer 3 consisting...

Страница 8: ...log to digital conversion is done at equal intervals of the optical path difference intervals The path difference reference scale is generated in a reference channel that consists of a HeNe laser the...

Страница 9: ...t 2 the Scan light 3 peripheral equipment connection ports 4 sample compartment 5 MCT detector compartment cover 6 spectrum recording control buttons Note The photo displays the device in basic config...

Страница 10: ...right sides The Power switch on the rear side of the device case near the 220 V power connector Device software installation First turn on the computer Run setup exe on the USB flash drive that comes...

Страница 11: ...r icons there indicate scan errors Single errors can be caused by the device being pushed or shaken If errors keep occurring or the LED on the case stops blinking and the software window indicator loo...

Страница 12: ...the beams reflected from an empty cell distort the spectrum Then click the Background Spectrum button or press Shift F9 The spectrum recording time is proportionate to the number of the scans and can...

Страница 13: ...triangle to the right of the button and select you add on from the list Note that the scan number resolution and other recording settings apply to the current selection so switching to another add on...

Страница 14: ...You can make any folder a home folder by opening it a clicking a context menu command Thus you can navigate the folder system on the disk through the list You can also open logical spectrum folders th...

Страница 15: ...ial if you edit it the file is renamed As the Number field has the line number entering a number in this field results in the line under this number selected e g you can use this feature to get a spec...

Страница 16: ...The Acceleration option involves all the CPU cores and multiplies the search speed almost by the number of the cores The Reset to initial values button restores the default values applicable to ordin...

Страница 17: ...cating the number of each of them or the range using a hyphen e g C17 N3 5 02 Cl note that you do not have to use 1 Add to your formula if you want it to include any other elements except the listed o...

Страница 18: ...ick the View button you can display them in advance Use the Compare button for comparison with the current spectrum and band coloring Use the Purge button to completely delete a peak list The From cur...

Страница 19: ...rve The last open spectrum is marked as number 1 the previously opened spectra are numbered in an ascending order by adding one Any curve can become number one if you click on it You can also change t...

Страница 20: ...ht Note that these variants do not affect the view if only one curve is open Sample spectra can be displayed both in the transmittance scale the quotient of two spectra current and background for each...

Страница 21: ...he dependence of the spectrum current ordinate on the abscissa and vice versa Current X and Y are marked with dashed lines Their values are specified in the text fields at the graph edges You can use...

Страница 22: ...it will change the base line or amplitude of the current spectrum in the area of the cursor moving If you magnify the curve images then your edits will be more accurate Plotting border lines involves...

Страница 23: ...luted multiple times and recorded again Continue diluting until the peak of the substance to be identify appears on the spectrum For greater accuracy we recommend that you repeat the entire procedure...

Страница 24: ...calculating concentration for that peak Consequently if a peak has graph a relevant concentration will show up near the peak height Creating and editing libraries As noted above spectrum libraries are...

Страница 25: ...rogram If it does not help unplug and plug the USB cable back in Next turn off the device wait for a minute turn it on again and restart the program If it does not help restart the computer 2 Fault sc...

Страница 26: ...rding settings set a different number where necessary If the channel number is correct check the sample compartment for foreign items or pollution 6 Fault software failure Manifestation the program do...

Страница 27: ...ery two years Use the Calibration command in the Recording menu or the toolbar button The procedure is performed in the Master mode and consists of five steps Note that you may take measurements in an...

Страница 28: ...equires a reference film and a comparison spectrum If you do not have them contact the spectrometer manufacturer or get a thin polystyrene layer ca 0 25 mm and record its spectrum and set the spectrum...

Страница 29: ...n measuring precision are assessed at the next step This measurement is the most time consuming and can take a few minutes Its results are entered in the parameter list At the final step click the Rea...

Страница 30: ...onditions are the same as the operating conditions Table 1 Place the spectrometer at least 10 cm away from the warehouse floor and walls and at least 50 cm from the heat sources Transportation and sto...

Страница 31: ...measuring attenuated total reflection and specular diffuse reflection spectra It consists of a hold down 1 FTIR and SDR flange seats the seat shown covered with a plug flange 2 in Figure 12 The plug...

Страница 32: ...remove the sample and clean the hold down tip and crystal surface with an alcohol or other solvent wetted cotton pad Liquid samples are placed at the FTIR element center in the form of a tiny drop do...

Страница 33: ......

Отзывы: