Figure 6. Peak (band) search conditions 

Search for Mixtures

 only takes into account the peaks in the comparison spectrum. If 

some of the query peaks are absent from the comparison spectrum, they do not affect the 

general evaluation. This works if your query refers to a mixture and contains many peaks from 

the different substance spectra. With the 

Mix

 option on, the spectra returned by the search 

will be added in pairs in a proportion that provides for the best approximation to the query 

spectrum. Such combinations are rated separately and added to search results in the form 

shown in Fig. 5. You can apply mixing to any search but it works better for 

Search for Mixtures

The following buttons are helpful with peak lists. Peak positions and widths are 

displayed in the curve field after search: green bands for matching peaks, blue ones for the 

others. If you click the 

View

 button, you can display them in advance. Use the 

Compare

 button 

for comparison with the current spectrum and band coloring. Use the 

Purge

  button to 

completely delete a peak list. The 

From curve

 button fills in the list again collecting peaks from 

the current black curve in accordance with the selected threshold height. If you want to select 

individual peaks only, move the cursor to the peak in your spectrum and when it is highlighted 

with a dashed line, left-click on it. The peak will be added to the list and marked with a vertical 

stripe on the image. Another click on it will delete the peak. Anyway, after supplementing your 

list you can correct it manually. 

If your peak table is typed or edited manually you should save it as a text file by clicking 

the 

To file

 button so that you can read it later by clicking the 

From file

 button. 

Use the 

Fragment

 tab to set structural formulas and their fragments for search; this is 

only used if structural formula libraries are available. Methods of entering structures can be 

found in the electronic manual.

 

 

Содержание 4434-801-59962935-08

Страница 1: ...IROS 05 IR Fourier Spectrometer Operating Manual FSVE 443430 005RE 2020 Ostec Instruments Moscow...

Страница 2: ...specialists who have read this manual are qualified to work on electrical devices up to 1 000 V and briefed on workplace safety rules The spectrometer involves operations with hazardous 220 V 50 Hz c...

Страница 3: ...n 6 Device software installation 6 First run 7 Spectrum recording 7 Spectrum reading and copying 9 Spectrum identification 12 Viewing spectra 15 Spectrum editing 18 Calibration 19 Creating and editing...

Страница 4: ......

Страница 5: ...434 801 59962935 08 Please see spectrometer operating conditions in Table 1 Spectrometer s electromagnetic interference resistance complies with GOST 30804 6 1 2013 and radiated interference with GOST...

Страница 6: ...ise ratio RMS in standard conditions in1 min at a resolution of 4 cm 1 in full bandwidth of 2000 2200 cm 1 40 000 Signal gain 1 2 4 8 Resolution 0 5 1 2 4 8 cm 1 Scanning speed spectra per minute For...

Страница 7: ...1 Spectrometer optical scheme The inlet optical system 2 converts the IR source 1 radiant energy into a convergent annular beam and makes it focus on the inlet port of the interferometer 3 consisting...

Страница 8: ...log to digital conversion is done at equal intervals of the optical path difference intervals The path difference reference scale is generated in a reference channel that consists of a HeNe laser the...

Страница 9: ...t 2 the Scan light 3 peripheral equipment connection ports 4 sample compartment 5 MCT detector compartment cover 6 spectrum recording control buttons Note The photo displays the device in basic config...

Страница 10: ...right sides The Power switch on the rear side of the device case near the 220 V power connector Device software installation First turn on the computer Run setup exe on the USB flash drive that comes...

Страница 11: ...r icons there indicate scan errors Single errors can be caused by the device being pushed or shaken If errors keep occurring or the LED on the case stops blinking and the software window indicator loo...

Страница 12: ...the beams reflected from an empty cell distort the spectrum Then click the Background Spectrum button or press Shift F9 The spectrum recording time is proportionate to the number of the scans and can...

Страница 13: ...triangle to the right of the button and select you add on from the list Note that the scan number resolution and other recording settings apply to the current selection so switching to another add on...

Страница 14: ...You can make any folder a home folder by opening it a clicking a context menu command Thus you can navigate the folder system on the disk through the list You can also open logical spectrum folders th...

Страница 15: ...ial if you edit it the file is renamed As the Number field has the line number entering a number in this field results in the line under this number selected e g you can use this feature to get a spec...

Страница 16: ...The Acceleration option involves all the CPU cores and multiplies the search speed almost by the number of the cores The Reset to initial values button restores the default values applicable to ordin...

Страница 17: ...cating the number of each of them or the range using a hyphen e g C17 N3 5 02 Cl note that you do not have to use 1 Add to your formula if you want it to include any other elements except the listed o...

Страница 18: ...ick the View button you can display them in advance Use the Compare button for comparison with the current spectrum and band coloring Use the Purge button to completely delete a peak list The From cur...

Страница 19: ...rve The last open spectrum is marked as number 1 the previously opened spectra are numbered in an ascending order by adding one Any curve can become number one if you click on it You can also change t...

Страница 20: ...ht Note that these variants do not affect the view if only one curve is open Sample spectra can be displayed both in the transmittance scale the quotient of two spectra current and background for each...

Страница 21: ...he dependence of the spectrum current ordinate on the abscissa and vice versa Current X and Y are marked with dashed lines Their values are specified in the text fields at the graph edges You can use...

Страница 22: ...it will change the base line or amplitude of the current spectrum in the area of the cursor moving If you magnify the curve images then your edits will be more accurate Plotting border lines involves...

Страница 23: ...luted multiple times and recorded again Continue diluting until the peak of the substance to be identify appears on the spectrum For greater accuracy we recommend that you repeat the entire procedure...

Страница 24: ...calculating concentration for that peak Consequently if a peak has graph a relevant concentration will show up near the peak height Creating and editing libraries As noted above spectrum libraries are...

Страница 25: ...rogram If it does not help unplug and plug the USB cable back in Next turn off the device wait for a minute turn it on again and restart the program If it does not help restart the computer 2 Fault sc...

Страница 26: ...rding settings set a different number where necessary If the channel number is correct check the sample compartment for foreign items or pollution 6 Fault software failure Manifestation the program do...

Страница 27: ...ery two years Use the Calibration command in the Recording menu or the toolbar button The procedure is performed in the Master mode and consists of five steps Note that you may take measurements in an...

Страница 28: ...equires a reference film and a comparison spectrum If you do not have them contact the spectrometer manufacturer or get a thin polystyrene layer ca 0 25 mm and record its spectrum and set the spectrum...

Страница 29: ...n measuring precision are assessed at the next step This measurement is the most time consuming and can take a few minutes Its results are entered in the parameter list At the final step click the Rea...

Страница 30: ...onditions are the same as the operating conditions Table 1 Place the spectrometer at least 10 cm away from the warehouse floor and walls and at least 50 cm from the heat sources Transportation and sto...

Страница 31: ...measuring attenuated total reflection and specular diffuse reflection spectra It consists of a hold down 1 FTIR and SDR flange seats the seat shown covered with a plug flange 2 in Figure 12 The plug...

Страница 32: ...remove the sample and clean the hold down tip and crystal surface with an alcohol or other solvent wetted cotton pad Liquid samples are placed at the FTIR element center in the form of a tiny drop do...

Страница 33: ......

Отзывы: