Ostec 4434-801-59962935-08 Скачать руководство пользователя страница 14

 
 
 
 
 
 
 
 
 

10 

 

Figure 4. ZaIR 3.5 software window 

Each list also has a 

Read on the fly

 option (the button with a butterfly icon). If it is on, 

the spectrum opens immediately after you select the line a closes when the selection shifts 

to another line. In this case, if you want to open two or more spectra, select them with the 

Ctrl

  (to select non-adjacent lines) or 

Shift

  (to select a continuous sequence of lines) key 

pressed. On the fly reading activates by default for search results, you can also use it for 

reading library contents. 

All key strokes go to one active list. The active list window is highlighted with a thin 

frame and a current line is marked with a dashed rectangle. Click on a window to activate it. 

The 

Tab

 key switches to the next list. In addition to spectra, the lists can contain folders and 

spectrum libraries. If you open a folder or library, the list will display the spectra in them. 

The 

One level up

 button takes you to the previous folder level. The 

Home

 button takes 

to the root folder of the particular list. You can make any folder a home folder by opening it 

a clicking a context menu command. 

Thus, you can navigate the folder system on the disk through the list. You can also open 

logical spectrum folders: these are search results and search libraries. To do so, use the 

Address

 text field above the list. This field has the full path to the currently open folder or 

library. If you enter a different address in this field and press Enter, you will open the folder 

at that address. You can display a drop-down folder list by clicking the button to the right of 

the 

Address

 field. The first item on this list is the current folder, then all computer drives, and 

then logical folders: search results and spectrum search libraries. The search results in the list 

are for the sake of completeness. Selecting them from the list does not usually make any 

sense as they open automatically after search, they do not have any results before that. 

To return to the current folder after viewing search results, select the topmost line in 

the drop-down list or click the 

One level up

 button. If a search library was open, 

One level up

 

will take you first to the search library list (this is another logical folder) and then to the 

current folder.

 

 

Содержание 4434-801-59962935-08

Страница 1: ...IROS 05 IR Fourier Spectrometer Operating Manual FSVE 443430 005RE 2020 Ostec Instruments Moscow...

Страница 2: ...specialists who have read this manual are qualified to work on electrical devices up to 1 000 V and briefed on workplace safety rules The spectrometer involves operations with hazardous 220 V 50 Hz c...

Страница 3: ...n 6 Device software installation 6 First run 7 Spectrum recording 7 Spectrum reading and copying 9 Spectrum identification 12 Viewing spectra 15 Spectrum editing 18 Calibration 19 Creating and editing...

Страница 4: ......

Страница 5: ...434 801 59962935 08 Please see spectrometer operating conditions in Table 1 Spectrometer s electromagnetic interference resistance complies with GOST 30804 6 1 2013 and radiated interference with GOST...

Страница 6: ...ise ratio RMS in standard conditions in1 min at a resolution of 4 cm 1 in full bandwidth of 2000 2200 cm 1 40 000 Signal gain 1 2 4 8 Resolution 0 5 1 2 4 8 cm 1 Scanning speed spectra per minute For...

Страница 7: ...1 Spectrometer optical scheme The inlet optical system 2 converts the IR source 1 radiant energy into a convergent annular beam and makes it focus on the inlet port of the interferometer 3 consisting...

Страница 8: ...log to digital conversion is done at equal intervals of the optical path difference intervals The path difference reference scale is generated in a reference channel that consists of a HeNe laser the...

Страница 9: ...t 2 the Scan light 3 peripheral equipment connection ports 4 sample compartment 5 MCT detector compartment cover 6 spectrum recording control buttons Note The photo displays the device in basic config...

Страница 10: ...right sides The Power switch on the rear side of the device case near the 220 V power connector Device software installation First turn on the computer Run setup exe on the USB flash drive that comes...

Страница 11: ...r icons there indicate scan errors Single errors can be caused by the device being pushed or shaken If errors keep occurring or the LED on the case stops blinking and the software window indicator loo...

Страница 12: ...the beams reflected from an empty cell distort the spectrum Then click the Background Spectrum button or press Shift F9 The spectrum recording time is proportionate to the number of the scans and can...

Страница 13: ...triangle to the right of the button and select you add on from the list Note that the scan number resolution and other recording settings apply to the current selection so switching to another add on...

Страница 14: ...You can make any folder a home folder by opening it a clicking a context menu command Thus you can navigate the folder system on the disk through the list You can also open logical spectrum folders th...

Страница 15: ...ial if you edit it the file is renamed As the Number field has the line number entering a number in this field results in the line under this number selected e g you can use this feature to get a spec...

Страница 16: ...The Acceleration option involves all the CPU cores and multiplies the search speed almost by the number of the cores The Reset to initial values button restores the default values applicable to ordin...

Страница 17: ...cating the number of each of them or the range using a hyphen e g C17 N3 5 02 Cl note that you do not have to use 1 Add to your formula if you want it to include any other elements except the listed o...

Страница 18: ...ick the View button you can display them in advance Use the Compare button for comparison with the current spectrum and band coloring Use the Purge button to completely delete a peak list The From cur...

Страница 19: ...rve The last open spectrum is marked as number 1 the previously opened spectra are numbered in an ascending order by adding one Any curve can become number one if you click on it You can also change t...

Страница 20: ...ht Note that these variants do not affect the view if only one curve is open Sample spectra can be displayed both in the transmittance scale the quotient of two spectra current and background for each...

Страница 21: ...he dependence of the spectrum current ordinate on the abscissa and vice versa Current X and Y are marked with dashed lines Their values are specified in the text fields at the graph edges You can use...

Страница 22: ...it will change the base line or amplitude of the current spectrum in the area of the cursor moving If you magnify the curve images then your edits will be more accurate Plotting border lines involves...

Страница 23: ...luted multiple times and recorded again Continue diluting until the peak of the substance to be identify appears on the spectrum For greater accuracy we recommend that you repeat the entire procedure...

Страница 24: ...calculating concentration for that peak Consequently if a peak has graph a relevant concentration will show up near the peak height Creating and editing libraries As noted above spectrum libraries are...

Страница 25: ...rogram If it does not help unplug and plug the USB cable back in Next turn off the device wait for a minute turn it on again and restart the program If it does not help restart the computer 2 Fault sc...

Страница 26: ...rding settings set a different number where necessary If the channel number is correct check the sample compartment for foreign items or pollution 6 Fault software failure Manifestation the program do...

Страница 27: ...ery two years Use the Calibration command in the Recording menu or the toolbar button The procedure is performed in the Master mode and consists of five steps Note that you may take measurements in an...

Страница 28: ...equires a reference film and a comparison spectrum If you do not have them contact the spectrometer manufacturer or get a thin polystyrene layer ca 0 25 mm and record its spectrum and set the spectrum...

Страница 29: ...n measuring precision are assessed at the next step This measurement is the most time consuming and can take a few minutes Its results are entered in the parameter list At the final step click the Rea...

Страница 30: ...onditions are the same as the operating conditions Table 1 Place the spectrometer at least 10 cm away from the warehouse floor and walls and at least 50 cm from the heat sources Transportation and sto...

Страница 31: ...measuring attenuated total reflection and specular diffuse reflection spectra It consists of a hold down 1 FTIR and SDR flange seats the seat shown covered with a plug flange 2 in Figure 12 The plug...

Страница 32: ...remove the sample and clean the hold down tip and crystal surface with an alcohol or other solvent wetted cotton pad Liquid samples are placed at the FTIR element center in the form of a tiny drop do...

Страница 33: ......

Отзывы: