6
MAINTAINANCE & CLEANING
Note
: The device should always be used with spectra socks and foot
covers to minimize dust and dirt from entering between carbon blades. It
is important to fix the spectra sock around pyramid location in order to
prevent noise issues coming from sand entering between the plates.
The device is weather-proof but not resistant to corrosion. Therefore, the
device should not come into contact with salt water or clorinated water.
The device is not designed to be used under extreme conditions like
diving or jumping into water. If for some reason the foot components get
wet parts should be dried with a lint free cloth. Should the device come
into contact with salt water, chlorinated water, dust or dirt it must be
rinsed with fresh water immediately.
Note:
The stiffness of the shock module can increase in temperatures
below -15 °C (+5 °F).
LIABILITY
The manufacturer recommends using the device only under the specified
conditions and for the intended purposes. The device must be
maintained according to the instructions for use. The manufacturer is
not liable for damage caused by component combinations that were not
authorized by the manufacturer.
COMPLIANCE
This component has been tested according to ISO 10328 standard to two
million load cycles. Depending on the amputee’s activity this
corresponds to a duration of use of two to three years. We recommend
carrying out regular yearly safety checks
ISO 10328 - “P” - “m”kg *)
*) Body mass limit not to be exceeded!
For specific conditions and limitations of use see
manufacturer’s written instructions on intended use!
In the standard mentioned,
test levels (P) are assigned to
a certain maximal body
masses (m in kg). In some
cases, which are marked with,
no test level is assigned to
the product related maximal
body mass. In these cases, the test loads have been adapted adequately
on the basis of the specified load level.
Category Össur
Category
Weight (Kg)
Lable text
1
52
ISO 10328 -
P3
52
kg
2
59
ISO 10328 -
P3
59
kg
3
68
ISO 10328 -
P3
68
kg
4
77
ISO 10328 -
P4
77
kg
5
88
ISO 10328 -
P4
88
kg
6
100
ISO 10328 -
P5
100
kg
7
116
ISO 10328 -
P6
116
kg
8
130
ISO 10328 -
P7
130
kg
• This manual is intended for use by a certified prosthetist.
Содержание RE-FLEX ROTATE
Страница 1: ...Instructions for Use RE FLEX ROTATE TM...
Страница 7: ...7 WARRANTY ssur provides a warranty of 36 months for this foot module and 2 months for the foot cover...
Страница 49: ...49 90 3 3 3b 3 4 Loctite 243 3 3 Loctite 243 5 6 15 20...
Страница 55: ...55 ssur...
Страница 56: ...56 1 2 243 15Nm 11ft lbs 1 3 4 4 2 1 3 3 8mm 90 90 2mm 180 4mm 270 6mm 360 8mm 90 2mm 180 4mm 270 6mm 360 8mm...