OPERATION

 

When the pressure reaches full inflation (within approximately 5 minutes) the 

normal pressure indicator will illuminate and the low pressure indicator will turn off.

 

Turn the pressure dial (see 7) to set a comfortable pressure level from 40-110kg 

according to the approximate weight of the patient.

 

Press the static button to activate static mode. The static indicator will illuminate. 

Static mode will commence in approximately 5 minutes. Press the static button 
again to switch back to alternating mode.

SYSTEM MAINTENANCE 

Maintenance at regular intervals (mandatory every 2 years) is necessary to preserve 
the function of the OSKA SeriesC-V2 alternating cushion system. 
Used filters must be exchanged within this period.

Maintenance/inspection should include checking whether the following is required:

 Replace membrane unit
 Replace air filter
 Replace synchronous motor
 Conduct function test/function check
 Sealing of case and attaching an inspector sticker
 Protective conductor check in accordance with DIN VDE 0751

3A

Содержание SeriesC-V2

Страница 1: ...oska uk com 0845 381 0600 1 SERIESC V2 USER INSTRUCTIONS...

Страница 2: ...2 oska uk com 0845 381 0600 CHANGING LIVES...

Страница 3: ...OBSERVE THE WARNING INSTRUCTIONS BEFORE USING THE SYSTEM CONTENTS The System Recommendations Capacity System Set up Operation System Maintenance Cleaning Technical Specification Troubleshooting Waste...

Страница 4: ...aturing a membrane pump and a cushion with 6 air cells arranged in a transverse manner The entire cell unit is encased in a cradle of visco foam The bielastic top cover and PVC outer base cover ensure...

Страница 5: ...pump see 3b ensuring that the tubes are not twisted or kinked Allow space for tubes to drop freely Switch on the pump power see 4 The power switch indicator should now illuminate and the pump will sta...

Страница 6: ...will commence in approximately 5 minutes Press the static button again to switch back to alternating mode SYSTEM MAINTENANCE Maintenance at regular intervals mandatory every 2 years is necessary to p...

Страница 7: ...facturers information as stated on the cleaning agent If in doubt you must request further information TECHNICAL SPECIFICATION PUMP OSKA SeriesC V2 Power supply 220 240V 50Hz 1A Air output 8 liter min...

Страница 8: ...pump is running but is not inflating the cushion The hoses are connected correctly Disconnect the hoses and re connect them in to the outlet connectors on the side of the pump Contact OSKA Service Th...

Страница 9: ...tors In the case of system failure please check Is there sufficient power supply Is the unit turned on Is the power cord connected properly and is the power plug inserted correctly Is the cushion conn...

Страница 10: ...ring inspection it is found that damages are due to wear and tear or are not subject to warranty then OSKA are empowered to claim the expenses inspection transport costs etc from the client The use of...

Страница 11: ...ect to assist the flames You must avoid smoking at all times in the immediate vicinity of this product Other hazards can include The use of petroleum based creams and gels such as Vaseline etc The man...

Страница 12: ...12 oska uk com 0845 381 0600 1 3 College Hill London EC4R 2RA oska uk com 0845 381 0600...

Отзывы: