Air Overlay Mattress User Guide
7
System maintenance
Maintenance at regular intervals (mandatory every
2 years) is necessary to preserve the function of
the OSKA Air Overlay alternating pressure system.
Used filters must be exchanged within this period.
Maintenance/inspection is carried out by the
producer or an authorised firm or engineer at the
client’s expense.
If the maintenance is not performed within the
first 2 years of the products life or not conducted
by an authorised firm or engineer then the
warranty claims become invalid. Damage or any
other loss of function arising from the lack of
maintenance or interferences by an unauthorised
party will also invalidate the product’s warranty.
Maintenance/inspection includes the following
procedures:
01. Replace membrane unit
02. Replace air filter
03. Replace synchronous motor
04. Conduct function test/function check
05. Sealing of case and attaching an inspection
sticker
06. Protective conductor check in accordance
with DIN VDE 0751
Cleaning
Dirt or stains should be removed using warm
water with a mild detergent and a cloth. Do not
use aggressive cleaning agents or solvents.
Regular wipe disinfection is recommended using
a suitable disinfectant cleaner and following
manufacturer’s guidelines.
The Top covers are machine washable. Care must
be taken, in order to maintain the integrity of the
covers.
Light Blue Cover – This cover can be machine
washed up to 70˚C for disinfection, following
manufacturer’s dosing directions. Tumble dry on
low heat and ensure cover is thoroughly dry
before replacing.
Dark Blue Cover – This cover can be machine
washed up to 95˚C for disinfection, following
manufacturers dosing directions. Tumble dry on
low heat and ensure cover is thoroughly dry
before replacing.
Technical specification
Pump
OSKA Air Overlay
Power supply:
220~240V/50Hz, 1A
Air output:
8 litre/min.
Pressure range:
30-60 mmHg
Cycle time:
12 min.
Static mode
Alarm reset mode
With normal/low pressure light
Size:
28cm(L) x 20cm(W) x 10cm(H)
Weight:
2.2 kg
Mattress
Size:
86.5cm(W) x 200cm(L) x 12.5cm(H)
Top cover type 1
– Light blue bi-elastic cover with
PU/Polyester coating and zipper.
Top cover type 2
– Dark blue bi-elastic cover with
PU/Polyester coating and zipper.
Air cells:
Nylon / PVC
Base:
Nylon / PVC
17 air cells
CPR valve for emergency procedures
Pillow function
Quick connectors
Содержание Air Overlay