Air Overlay Mattress User Guide
5
System set-up
01. Place pump on horizontal surface or hang by
means of the hook.
02. Open mattress on the bed, secure it with the
end flaps or straps. Air tubes must be at foot
end of bed.
03. Connect air feed tubes to air outlet on the
mattress and pump (see 1), ensuring that the
tubes are not kinked or twisted. Allow space
for bed rails to drop freely.
04. Turn on the pump power (see 2). The power
switch indicator will come on. The pump will
start pumping air into the mattress.
05. The low pressure indicator (see 3 yellow or
red) will come on as inflation takes place.
06. The audible alarm won’t alarm within 60
minutes and will be activated if the pressure
is still low inside the mattress. In case of low
pressure, check the connection and products.
Press the alarm reset button (see 4) to mute
the audible alarm, it’s indicator will flash.
07. The pump will automatically default to setting
‘5’ whilst inflating. Leave the mattress to fully
inflate for 60 minutes before attempting to
change the pump setting.
1
4
5
3
2
Operation
01. When the pressure reaches the pre-set
level (within approximately 30 minutes), the
normal pressure indicator will come on and
the low pressure indicator will turn off.
02. Turn the pressure dial (see 5) to set a
comfortable pressure level from 40kg –
130kg according to the approximate weight
of the patient.
03. Press the static button to activate static
mode, the static indicator will come on. The
static mode will start within approximately
6 minutes. Press the static button again to
switch back to alternating mode.
Bi elastic vapour
permeable cover
Manual pump
CPR valve
Anchor end
straps