background image

ABOUT YOUR MATTRESS

HOW TO SET UP YOUR

MATTRESS

COVER:

The Series3-V4 comes with a
cover which provide an
impenetrable barrier to viruses and
bacteria. The cover has welded seams
that are ingress proof, minimising any
risk of harbouring viruses or bacteria
and transmitting infections.

The cover includes a fire evacuation
base which means handles can be easily
accessed in case of emergency and
used for moving and transporting.

An additional inner cover protects the
foam and cells.

THE SYSTEM

OSKA Series3-V4 alternating pressure 
system is a replacement system which is 
designed to be placed directly onto the 
support surface of a nursing bed.

Application in regular bed types is pos-
sible, however, preference is given to a 
patient profile nursing bed.

The system consists of control unit 
featuring a membrane pump and a bed 
overlay with 20 air cells arranged in a 
transverse manner. The entire cell unit is 
equipped with a bi-elastic PU protection 
cover. 

RECOMMENDATIONS:

OSKA

TM

 Series3-V4 self adjusting

replacement pressure system is
recommended to be used as an aid
in the prevention and treatment of
pressure ulcers up-to and including
category 4.

CAPACITY:

OSKA

TM

 Series3-V4 is recommended for

UNPACKING THE MATTRESS

Please ensure the box is not damaged
upon arrival. Avoid using a knife
when opening the box and any other
protective material as this could damage
the mattress and/or cover.

Place the Series3-V4 onto the bed frame
and adjust the cover if needed. The
mattress has a black side that should
be down on the bed frame, and a blue
fabric side that should face up towards
the patient.

HAND CHECK

The hand check should be conducted
regularly to check the mattress is
functioning correctly.

With the cover open, using a vertical
hand, insert your hand between the
air and cells directly underneath the
patient’s sacrum. The correct clearance
between the mattress base and patient
is 2-3 fingers. Check the cells and tubing
if this is not the case. If problem persists
please contact OSKA

TM

.

Confirm that the bed frame is
appropriate for use with the mattress,
and that the length and width of the
mattress are appropriate for the frame.

Place directly on a healthcare bedframe
only, never on top of another mattress.
Leave for a few minutes before using.

patient weight up to 175kg. Heavier
patients are better accommodated with
a more appropriate OSKA

TM

 pressure

system.

You must not lift the mattress whilst
a patient is lying on it or use it to
transport a patient unless in an
evacuation.

Содержание 3-V4 Series

Страница 1: ...of shear damage The easy to use fully digital pump has very clear weight settings and alternating cycle settings h Risk y 4 ment mfort control to wires gainst t of t WELDED SEAMS Regular sewn seams ar...

Страница 2: ...2 oska uk com 0845 381 0600 CHANGING LIVESTM...

Страница 3: ...E THE WARNING INSTRUCTIONS BEFORE USING THE SYSTEM CONTENTS Know your symbols Warning About your mattress How to set up your mattress System set up Cleaning Safety General information Warranty Trouble...

Страница 4: ...at persons must obey to avoid equipment damage Machine wash at 95 Manufacturer Drip dry Follow Instructions for Use 1 Chlorine Foot end Do not iron Group of function Tumble dry Separate Collection use...

Страница 5: ...imultaneously DO NOT USE Position the mattress correctly to prevent it becoming trapped and damaged NOTE Bed rails can trap the mattress and cause damage length and width of the mattress are appropria...

Страница 6: ...y 4 CAPACITY OSKATM Series3 V4 is recommended for UNPACKING THE MATTRESS Please ensure the box is not damaged upon arrival Avoid using a knife when opening the box and any other protective material as...

Страница 7: ...ton and its indicator will come on see 6 Press the button again to switch back to alternate mode OSKA 2016 4 1 2 3 4 5 6 Place pump on horizontal surface or hang by means of the hook Open mattress on...

Страница 8: ...inutes In the case of a poor connection with tubing or any other air leakage from the mattress low pressure conditions will occur and the audible visual alarm will be activated OPERATION 1 During norm...

Страница 9: ...lest breach of this barrier will cause fluid to penetrate the mattress There are specific cleaning and care instructions that need to be strictly adhered to prolong the life of the product DELAM Delam...

Страница 10: ...nt They may be disposed of with the household waste The pump control unit is considered as electrical equipment waste Please contact the proper waste disposal officials Part Description Part No In the...

Страница 11: ...uring cleaning extra care should be taken to avoid abrasion of the surface OSKATM covers incorporate a special 1 Always abide by the cleaning instruc tions 2 Avoid using chemicals with too high concen...

Страница 12: ...subject to regular service and maintenance a warranty can be offered only if the service intervals prescribed by OSKATM are adhered to If during inspection it is found that damages are due to wear and...

Страница 13: ...et connectors on the side of the pump Contact OSKA Service There are no leaks in the mattress Contact OSKA Service Contact OSKA Service There are no kinks in the tubing down side of the mattress Caref...

Страница 14: ...nance or interferences by an unauthorised party will also invalidate the product s warranty OSKA Series3 V4 Power supply 220 240V 50Hz 1A Air output 7 liter min Pressure range 20 70 mmHg Cycle time 10...

Страница 15: ...know that every patients need is different and so do we That s why our 3c solution helps us to help you deliver the highest standard of care possible CLINICAL Outstanding performance Prevention is be...

Страница 16: ...1 3 College Hill London EC4R 2RA 2 Selgar Ave Clovelly Park SA 5042...

Отзывы: