12
Storage
When moving the SpinalStim from very cold or very hot storage areas (like
your car), wait at least an hour to use or charge the device. The device
requires time to return to operating temperature.
Unpacked Storage:
Temperature Range:
• -25°C to 5°C
• 5°C to 35°C at up to 90% relative humidity, non-condensing
• 35°C to 60°C at a water vapor pressure up to 50 hPa
Packed Storage, Shipping and Transport:
Temperature Range: within -40°C to 60°C
• Between 10-100% relative humidity
• Including condensation at pressures between 500 hPa and 1060 hPa
Operating Environment:
Temperature Range: within 5°C to 40°C
• 15-90% relative humidity, non-condensing but not requiring a water
vapor pressure greater than 50 hPa
• 700-1060 hPa
SpinalStim is designed for a storage life of twelve months plus one year
of usage.
• DO NOT expose SpinalStim to direct sunlight for long periods of time.
• DO NOT expose SpinalStim to excessive heat or cold.
• Avoid storing the device in areas prone to extreme temperatures, such as
an enclosed automobile or trunk.
Travel
When traveling by air, it is recommended to pack SpinalStim with checked
luggage. If taken onboard the airplane, it should be turned off when passing
through security screening equipment, as the device could be damaged. The
SpinalStim instruction manual should be taken with you to quickly and easily
identify the device for security personnel. Do not wear or operate SpinalStim
while onboard the airplane.