5.16 System Information
radEAGLE
User Manual
5.16 System Information
This screen (Fig.
, p.
) shows information about your
radEAGLE
’s hardware status.
Please have these data at hand when contacting our support.
Figure 71:
The hardware status
High Voltage
Typical values are between 500 V and 800 V.
Battery Temperature
The value should be between 0 °C and 50 °C. (Available only only for the black
power-
CELL
, see
, p.
).
Crystal Temperature
The value should be between -20 °C and 55 °C.
Fine Gain
The value should be ±10 %.
Last Stabilization
Time elapsed since the last successfull stabilization of the instrument (see
p.
Exit
Return to the superior screen.
5.17 Alarm Settings
This screen (Fig.
, p.
) groups the settings for warnings and alarms.
These settings are designed for your personal safety. The alarm is intended to let you
know you are exposed to dangerous radiation and may be accumulating a significant
radiation dose. If you set the alarm levels or warning thresholds too high, this may
pose a serious health risk to you.
You can set the unit used for these thresholds under
, p.
.
66/147
Software 3.2.12 • Document 3.3.0o
© innoRIID GmbH • 2019-04-11
Содержание radEAGLE
Страница 1: ...radEAGLE User Manual Software 3 2 12 Document 3 3 0o 2019 04 11 ...
Страница 20: ...20 147 Software 3 2 12 Document 3 3 0o innoRIID GmbH 2019 04 11 ...
Страница 40: ...40 147 Software 3 2 12 Document 3 3 0o innoRIID GmbH 2019 04 11 ...
Страница 106: ...106 147 Software 3 2 12 Document 3 3 0o innoRIID GmbH 2019 04 11 ...
Страница 118: ...118 147 Software 3 2 12 Document 3 3 0o innoRIID GmbH 2019 04 11 ...
Страница 134: ...134 147 Software 3 2 12 Document 3 3 0o innoRIID GmbH 2019 04 11 ...