radEAGLE
User Manual
radEAGLE
Advanced Menu
Figure 72:
The alarm settings
Warning Threshold
The dose rate threshold for warnings. The value must be lower than that for alarms.
Alarm Threshold
The dose rate threshold for alarms. The value must be higher than that for warnings.
Neutron Sensitivity
Set sensitivity for neutron alarms in several steps.
This item is available for
radEAGLE
models with a neutron detector (see
p.
).
Setting
Neutrons
Integration Duration
-5
4
4 sec
-4
4
10 sec
-3
4
15 sec
-2
4
20 sec
-1
4
30 sec
Default ±0
4
40 sec
+1
4
80 sec
+2
4
100 sec
+3
4
140 sec
+4
4
180 sec
+5
4
220 sec
Exit
Return to the superior screen.
Down
Highlight the next item. Cycles to the top item after you reached the last item.
Change
Edit the highlighted setting.
© innoRIID GmbH • 2019-04-11
Software 3.2.12 • Document 3.3.0o
67/147