8. TROUBLESHOOTING
Error Message
Meaning
Action
61
The Cooler Is Turned Off
(as stated)
Turn cooler on (Section 3.3.2) and allow
at least 12 hours for detector to fully cool.
Monitor status according to
Section 3.4.2.1.
Internal System Error
Internal error
Contact ORTEC.
Error Retrieving Health
Internal error
Contact ORTEC.
Error Retrieving Data
Internal error
Contact ORTEC.
Bias Voltage Error
Bias voltage is outside of acceptable
limits
Contact ORTEC.
Low Battery!
Available battery time is critically
low
Connect to external power. If battery
exhausts during a survey, the current
survey data set could be lost.
Storage is Low!
Data card is nearly full
End the current survey, return to Home
screen, and insert memory card with
available storage space; or transfer files to
PC (Chapter 7), then use Clear Data
command to permanently delete all data
files from SD card (Section A.4).
Data Card Full/Missing!
Data card is full or has been
removed
From Home screen, insert memory card
with available storage space. If the amber
indicator does not clear, see Section 8.7.
+12 Volt Error
+12 V power supply voltage is
outside of acceptable limits
Contact ORTEC.
-12 Volt Error
!
12 V power supply voltage is
outside of acceptable limits
Contact ORTEC.
+3 Volt Error
+3 V power supply voltage is
outside of acceptable limits
Contact ORTEC.
Detector Temp High
Detector temperature is too high or
detector cold-tip temperature is high
Make sure cooler is on (Section 3.3.2),
allow at least 12 hours for detector to
fully cool. Monitor detailed status
according to Section 3.4.2.1. If detector
does not cool, contact ORTEC.
Ion Pump Error
Ion pump voltage is outside of
acceptable limits
Contact ORTEC.
New Background Required
Background spectrum is out of date
Perform background check
(Section 3.3.4).
Содержание Micro-Detective-HX
Страница 4: ...iv...
Страница 10: ...x...
Страница 26: ...ORTEC Micro Detective HX Human Portable Radiation Detection System Operation Manual 16...
Страница 45: ...3 GETTING STARTED 35 Fig 30 HX Program Menu Structure Administrator Access Password Required...
Страница 46: ...ORTEC Micro Detective HX Human Portable Radiation Detection System Operation Manual 36...
Страница 62: ...ORTEC Micro Detective HX Human Portable Radiation Detection System Operation Manual 52...
Страница 68: ...ORTEC Micro Detective HX Human Portable Radiation Detection System Operation Manual 58...
Страница 74: ...ORTEC Micro Detective HX Human Portable Radiation Detection System Operation Manual 64...
Страница 82: ...ORTEC Micro Detective HX Human Portable Radiation Detection System Operation Manual 72...
Страница 92: ...ORTEC Micro Detective HX Human Portable Radiation Detection System Operation Manual 82...