The inputs/outputs of the SUB-D connector are defined in the following table:
GND
Ground
CTRL IN & OUT
Can be used to connect a Serial Terminal port to manage Network configuration
without Network RJ45 port.
RL_URG_C
Urgent Alarm
–
Common contact
RL_URG_T
Urgent Alarm
–
Active opened contact
RL_URG_R
Urgent Alarm
–
Active closed contact
RL_NON-URG_C
Non-Urgent Alarm
–
Common contact
RL_NON-URG_T
Non-Urgent Alarm
–
Active opened contact
RL_NON-URG_R
Non-Urgent Alarm
–
Active closed contact
To get a closed contact when an Urgent Alarm (see 11.2.1) is raised, 1 and 7 pins shall be used.
To get a closed contact when a Non-Urgent Alarm (see 11.2.2) is raised, 4 and 8 pins shall be used.
To get an opened contact when an Urgent Alarm (see 11.2.1) is raised, 1 and 6 pins shall be used.
To get an opened contact when a Non-Urgent Alarm (see 11.2.2) is raised, 4 and 9 pins shall be used.
Additional RS232 Serial port function (No CLI access):
Accepts commands to locally configure the IP network
parameters for initial connectivity - Character structure: ASCII, 9600 baud, 1 start, 8 data, 1 stop, no parity. Please,
use only a dedicated cable with Pin 2,3 and 5 to use this feature independently
–
separated from Alarm contacts.
Содержание Spectracom EC20S
Страница 44: ...EPSILON CLOCK MODEL EC20S Spectracom User s Manual 44 13 3 CLOCK STATUS PAGE...
Страница 58: ...EPSILON CLOCK MODEL EC20S Spectracom User s Manual 58 13 6 SNMP TRAPS SETUP PAGE...
Страница 60: ...EPSILON CLOCK MODEL EC20S Spectracom User s Manual 60 No No trap is generated...
Страница 82: ...EPSILON CLOCK MODEL EC20S Spectracom User s Manual 82 13 13 ADMIN PAGE...
Страница 91: ...EPSILON CLOCK MODEL EC20S Spectracom 91 User s Manual 98 RESTART EC20S 1 the EC20S unit restarts...