O2k-FluoRespirometer manual
7
Oroboros Instruments
High-Resolution FluoRespirometry
5.3. O2k OroboPOS assembly
The POS are shipped dry without electrolyte in
a Perspex housing in the OroboPOS-Service Kit.
Before assembling the O2k, apply electrolyte
and membrane, following the OroboPOS service
instructions.
Go to: OroboPOS assembly
»
MiPNet19.18B_POS-service
1
. Screw a POS head onto each
POS connector
(finger-tight). Push a
wetted
POS seal tip (black gasket)
over the POS head (arrow 1). Position the pore centrally
and do not stretch the gasket.
2
. Connect the POS connector cable to the O2k-Main
Unit. Insert the male plug of the cable 2 into the female
O2 plug. The red dot on the male plug 2 faces straight
upwards. Each OroboPOS head can be used on O2k-
chamber A or B. Note the POS number (marked on
the cylindrical body of each POS) for each chamber
(see
Section 6
)
.
3
. After inserting the plug of the POS
connector cable to the O2k-Main unit,
the POS connector cable 3 must be in a
relaxed, undistorted position. Press the
sleeve downwards (arrow 4), insert the
POS with the black gasket straight into the POS
holder 5, and release the sleeve for fixing the POS
in its final position.
Do not rotate the POS connector
6 after it
is attached to the POS holder in order to prevent
damage of the black gaskets. The gaskets provide
a tight seal against the glass chamber that has
sharp edges.
Clip A2
2
1
2
4
5
3
6