MiPNet22.11
O2k-FluoRespirometer manual
5
Oroboros Instruments
High-Resolution FluoRespirometry
Note: Use O2k-Main Power Cable\230 V\Europe (Hybrid
German/French plug CEE7/7, Type E+F) only for a
compatible socket.
DK:
Some Danish sockets will also accept the CEE7/7 plugs,
but there is no grounding connection with these plugs
without male ground pin. Exchange for an appropriate
Danish power cable.
CH:
A different main power cable or adapter is required.
US:
O2k-Main Power Cable\120 V\US-CA, 21112-01
AU/NZ:
O2k-Main Power Cable\230 V\AU-NZ, 21113-01
The
O2k
requires a minimum of lab space (0.5 m bench
space) next to an external PC or laptop, with sufficient
space for the
ISS
,
O2k-Titration Set
and handling. For
maximum signal stability and optimal temperature
control, the O2k should be placed in a position in which
it is not exposed to direct sunlight and sudden changes
of room temperature. Keep the space on the table below
the O2k bottom clear to prevent any interference of
ventilation beneath the
O2k-Main Unit
.
Avoid damage due to condensation. When placing the
O2k from a cold to warm environment, let it equilibrate
for at least 4 h in the warm room before connecting to
the mains.
Rear of the O2k-FluoRespirometer.
New features are the USB hub 2.0 with 4 external ports,
a plug for an external Pt1000 thermistor, and an optional
extension through an RS485 serial port.
WGT quality control seal:
Do not open. An intact quality control seal
is required for warranty obligations related to the
electronics of the O2k.