Oroboros Instruments O2k-Core Скачать руководство пользователя страница 6

MiPNet19.18 

B: OroboPOS service

 

22

 

 

 

OROBOROS INSTRUMENTS

 

 

O2k – tested and trusted 

evaporation of ethanol. Before an experiment, the ethanol is siphoned off 

and the chamber is washed with distilled water (

MiPNet06.03

)

 

5.2.  Short-term shelf storage

 

 

For  shelf  storage,  unplug  the  POS  connector  from  the  O2k-Main  Unit. 
Clean the sealing tip and membrane with distilled water. Mount the POS to 

the  blue  storage  base  and  seal  it  with  the  Perspex  cup  to 

prevent  the 

drying out of the electrolyte

. Store in the dark. 

 

5.3.  Long-term storage

 

 

For storage of the POS for several months, the sealing tip and membrane 
are removed. Wash the electrolyte off the POS with distilled water. Even if 
the  membrane  is  not  damaged,  remove  it  by  gripping  the  membrane 

holding ring with the groove in the lower end of the membrane ring holder 
of the OroboPOS-Mounting Tool (see above). The POS head is 

stored dry 

and  in  the  dark

.  Check  for  any  moisture  and  salt  contamination  in  the 

electrical connector of the POS head. In case, wash with distilled water and 

subsequently with pure methanol, dry at 60 

o

C for 24 h. 

 

6.

 

Replacement of the OroboPOS head

 

 

After  cleaning  and  drying  the  gold  cathode  contact  or  drying  the  sensor 

body  at  60  °C  for  a  few  days,  the  zero  current  should  be  reduced.  In 
addition,  the  zero  current  of  the  bare  cable,  without  sensor  head 

connected, is tested for any leak currents. If the latter test excludes any 
sources  of  leak  currents  other  than  the  POS,  and  POS  service  is  not 

successful, the sensor head must be replaced. 

 

 

A  new  sensor  head  can  be  screwed  onto  the  sensor  connector,  if  the 

old sensor head has been irreversibly damaged or should be replaced. 

 

7.

 

References

 

 

 

Gnaiger  E  (2008)  Polarographic  oxygen  sensors,  the  oxygraph  and 
high-resolution  respirometry  to  assess  mitochondrial  function.  In: 
Mitochondrial  dysfunction  in  drug-induced  toxicity  (Dykens  JA,  Will  Y 
eds) John Wiley:327-52.

 

»

 

 

Gnaiger  E,  Forstner  H,  eds  (1983)  Polarographic  Oxygen  Sensors. 
Aquatic  and  Physiological  Applications.  Springer,  Berlin,  Heidelberg, 
New York:370 pp. »

 

 

Hitchman  ML  (1983)  Calibration  and  accuracy  of  polarographic  oxygen 
sensors.  In:  Polarographic  oxygen  sensors,  Gnaiger  E  and  Forstner  H, 
eds., Springer, Berlin, Heidelberg, New York:18-30. »

 

 

 

Hitchman  ML,  Gnaiger  E  (1983)  A  thermodynamic  consideration  of 
permeability  coefficients  of  membranes.  In:  Polarographic  oxygen 
sensors,  Gnaiger  E  and  Forstner  H,  eds.,  Springer,  Berlin,  Heidelberg, 
New York:31-6. »

 

 

Updates »

http://wiki.oroboros.at/index.php/MiPNet19.18B_POS-Service

 

 

»

http://wiki.oroboros.at/index.php/MiPNet19.18A_O2k-Start

 

 

Next step

 – O2k-Core Manual D

 

»

MiPNet19.18D O2k-Calibration

 

Содержание O2k-Core

Страница 1: ...Videosupport Summary Service of the polarographic oxygen sensors OroboPOS POS is the basis of signal stability low noise and high time resolution Performance specifications of the OROBOROS O2k can be...

Страница 2: ...ensor service are provided in the OroboPOS Service Kit In addition distilled water and 25 ammonia solution fresh are required Store the OroboPOS in the dark 2 Cleaning Prevent damage by electrostatic...

Страница 3: ...or one minute in a figure eight motion Wash the polishing powder carefully off the sensor with distilled water Subsequently the anode is always cleaned as well Section 2 2 If noise remains high or the...

Страница 4: ...me of 20 ml Store at room temperature or 4 C in a closed vial To prevent the formation of gas bubbles in the electrolyte heat the electrolyte solution by shaking the electrolyte vial in hot water 40 7...

Страница 5: ...le connector next to the control light of the stirrer O2k Main Unit The red dot on the male plug has to face accurately upwards when inserting the plug Avoid bending and torsions of the cable MiPNet19...

Страница 6: ...ody at 60 C for a few days the zero current should be reduced In addition the zero current of the bare cable without sensor head connected is tested for any leak currents If the latter test excludes a...

Страница 7: ...ROBOROS USB flash drive B1_OroboPOS from Perspex housing B2 Disassembly of OroboPOS B3 Cathode cleaning OroboPOS Polishing Cloth OroboPOS Polishing Powder B4 Anode cleaning B5 Membrane mounting Supple...

Отзывы: