MiPNet19.18 

B: OroboPOS service

 

20

 

 

 

OROBOROS INSTRUMENTS

 

 

O2k – tested and trusted 

2.3.  Clean the electrical connection

 

 

Unscrew  the POS head and  inspect  both sides of  the electrical connection 

(gold  pin  and  threads).  Remove  any  contamination  such  as  salt  crystals, 
grease  and  moisture  with  a  fine  paper  cloth.  If  necessary,  wash  with 
distilled water and then with pure alcohol. Appl

Pen-Contact Oil

 from the 

OroboPOS-Service Kit to the connections only if necessary. Before screwing 
the  POS  head  onto  the 

OroboPOS-Connector

  for  membrane  application, 

clean  the  POS  connector  of  moisture  and  any  other  contamination 
(particularly any salt crystals from the electrolyte). Similarly, clean the plug 

of the electrical cable connecting to the O2k-Main Unit. 

 

3.

 

Membrane mounting

 

 

The 

POS head

 is screwed onto the blue 

OroboPOS-Connector

. Check the O-

ring  on  the  POS  head  to  ensure  that  its  surface  is  smooth  and  intact.  In 
exceptional  cases,  apply  a  tiny  amount  of  grease  to  the  O-ring  of  the 

sensor head. 

 

 

Use  KCl  solution  as  electrolyte  (1  mol∙dm

-3

;  74.56  g  potassium 

chloride per litre, in distilled water) provided in the OroboPOS-Service Kit. 
Add  distilled  water  to  the 

electrolyte  powder

  up  to  the  10  ml  mark. 

Alternatively, dissolve 1.49 g KCl in distilled water with a total volume of 20 
ml.  Store  at  room  temperature  or  4  °C  in  a  closed  vial.  To  prevent  the 

formation of gas bubbles in the electrolyte, heat the electrolyte solution by 
shaking  the  electrolyte  vial  in  hot  water  (40-70  °C)  before  membrane 

mounting. 

 

 

POS membranes

 are contained in small boxes in the OroboPOS-Service 

Kit.  They  are  fully  transparent.  Each  membrane  is  separated  by  a  non-

transparent white paper sheet. Do not add the paper to the oxygen sensor. 
Separate  a  membrane  from  the  stack  of  paper  sheets  and  membranes, 

carefully  avoiding  any  mechanical  damage  of  the  transparent  membrane. 
Do not touch the central area of the membrane with your fingers. 

 

 

  The  black 

OroboPOS-Mounting  Tool

  consists  of  two  parts,  (i)  the 

membrane guide (larger diameter) and (ii) membrane ring holder with O-

ring. Wash off any electrolyte and salt crystals with distilled water. 

 
 

 

1 Position a new membrane into the 

membrane guide

 using the 

pair 

of  forceps

.  2  Fix  the 

OroboPOS-Membrane  Ring

  (which  seales  the 

membrane  against  the  sensor  body)  to  the 

membrane  ring  holder

.  Fill 

the  POS  head  with  electrolyte.  3  Slide  the 

membrane  guide

  downwards 

across  the  POS  head  while  pushing  the  lower  ring  (arrow)  of  the  blue 

OroboPOS-Connector

  strongly  downwards.  By  releasing  this  ring,  the 

membrane guide is fixed to the 

POS connector

.  4 To slide the OroboPOS-

Membrane  Ring  over  the  POS  head,  slide  the  membrane  ring  holder  into 

the membrane guide, and press firmly down in a single movement to the 
final position. 

Clip B5

 

Содержание O2k-Core

Страница 1: ...Videosupport Summary Service of the polarographic oxygen sensors OroboPOS POS is the basis of signal stability low noise and high time resolution Performance specifications of the OROBOROS O2k can be...

Страница 2: ...ensor service are provided in the OroboPOS Service Kit In addition distilled water and 25 ammonia solution fresh are required Store the OroboPOS in the dark 2 Cleaning Prevent damage by electrostatic...

Страница 3: ...or one minute in a figure eight motion Wash the polishing powder carefully off the sensor with distilled water Subsequently the anode is always cleaned as well Section 2 2 If noise remains high or the...

Страница 4: ...me of 20 ml Store at room temperature or 4 C in a closed vial To prevent the formation of gas bubbles in the electrolyte heat the electrolyte solution by shaking the electrolyte vial in hot water 40 7...

Страница 5: ...le connector next to the control light of the stirrer O2k Main Unit The red dot on the male plug has to face accurately upwards when inserting the plug Avoid bending and torsions of the cable MiPNet19...

Страница 6: ...ody at 60 C for a few days the zero current should be reduced In addition the zero current of the bare cable without sensor head connected is tested for any leak currents If the latter test excludes a...

Страница 7: ...ROBOROS USB flash drive B1_OroboPOS from Perspex housing B2 Disassembly of OroboPOS B3 Cathode cleaning OroboPOS Polishing Cloth OroboPOS Polishing Powder B4 Anode cleaning B5 Membrane mounting Supple...

Отзывы: