B: OroboPOS service

 

21

 

 

 

OROBOROS INSTRUMENTS

 

 

high-resolution respirometry

 

 

• 

Mounting a membrane onto the OroboPOS

, using 

the  OroboPOS-Mounting  Tool  which  consists  of  two 

parts,  the  membrane  guide  and  the  membrane  ring 
holder. 

No  gas  bubbles 

should be trapped 
in  the  electrolyte 

reservoir 

after 

membrane 

application. 

No 

folds  should  be 

visible 

in 

the 

membrane  in  the 

central 

area. 

Inspect 

the 

electrolyte 

reservoir  under  a 
binocular.  Small 

folds 

in 

the 

membrane  near 

the 

outer 

circumference 

have 

no 

influence, but large folds should be avoided. Wash excess electrolyte off the 

POS  and  POS  connector.  Apply  a  wet 

OroboPOS-Seal  Tip

  and  attach  the 

OroboPOS-Connector

 to the 

OroboPOS-Holder

 (O2k-Main Unit, Chapter A). 

 

4.

 

Electrical cable connection 

 

Insert  the  male  plug  of  the  cable  of  the 

OroboPOS-Connector

  into  the  female  connector 

next to the control  light of the stirrer (O2k-Main 

Unit).  The  red  dot  on  the  male  plug  has  to  face 
accurately  upwards  when  inserting  the  plug. 

Avoid bending and torsions of the cable. 

 

» MiPNet19.01 for O2k Series B to C.

 

 

 

After  sensor  service  and  membrane  mounting,  the  POS  needs  some 

time  in  operation  to  stabilize  while  the  O2k  is  switched  on.  Such 

stabilization  may  require  several  hours.  For  this  purpose  the  O2k  may  be 
switched on and run overnight. 

 

5.

 

Storage of the OroboPOS

 

 

5.1.  Short-term storage in the O2k

 

 

For short periods of days or several weeks, the POS is maintained in the 
O2k chamber. The chamber is washed with distilled water and completely 

filled  with  70%  ethanol  for  chemical  sterilization.  The  stopper  is  inserted 
loosely without pushing it down beyond the point where the sealing ring is 

inserted  into  the  glass  chamber.  This  ensures  a  longer  life  time  of  the 
sealing  rings.  The  receptacle  of  the  stopper  is  completely  filled  with 

ethanol  from  the  top,  and  is  sealed  with  a  black  cover  slip  to  avoid 

Содержание O2k-Core

Страница 1: ...Videosupport Summary Service of the polarographic oxygen sensors OroboPOS POS is the basis of signal stability low noise and high time resolution Performance specifications of the OROBOROS O2k can be...

Страница 2: ...ensor service are provided in the OroboPOS Service Kit In addition distilled water and 25 ammonia solution fresh are required Store the OroboPOS in the dark 2 Cleaning Prevent damage by electrostatic...

Страница 3: ...or one minute in a figure eight motion Wash the polishing powder carefully off the sensor with distilled water Subsequently the anode is always cleaned as well Section 2 2 If noise remains high or the...

Страница 4: ...me of 20 ml Store at room temperature or 4 C in a closed vial To prevent the formation of gas bubbles in the electrolyte heat the electrolyte solution by shaking the electrolyte vial in hot water 40 7...

Страница 5: ...le connector next to the control light of the stirrer O2k Main Unit The red dot on the male plug has to face accurately upwards when inserting the plug Avoid bending and torsions of the cable MiPNet19...

Страница 6: ...ody at 60 C for a few days the zero current should be reduced In addition the zero current of the bare cable without sensor head connected is tested for any leak currents If the latter test excludes a...

Страница 7: ...ROBOROS USB flash drive B1_OroboPOS from Perspex housing B2 Disassembly of OroboPOS B3 Cathode cleaning OroboPOS Polishing Cloth OroboPOS Polishing Powder B4 Anode cleaning B5 Membrane mounting Supple...

Отзывы: