5 Motor over current:

The current for the motor exceeded the maximum limit

-> 

Check the mechanics (Pos. 30)

-> 

Replace motor control (Pos. 25)

-> 

Replace the motor (Pos. 26)

6 Motor control error:

   Internal error in the motor control.

-> 

Replace motor control (Pos. 25)

7 Eprom access error:

   Memory error in the access of the EPROM

-> 

Replace the hand held programming unit (Pos. 20)

8 CPM ROM error:

   Memory error in the motor control

-> 

Replace motor control (Pos. 25)

9 Communication:

   Communication to the motor control is not possible

-> 

Check spiral cable and connector

-> 

Replace the hand held programming unit (Pos. 20)

-> 

Replace motor control (Pos. 25)

10 General error in the motor control:

     Unknown error in the motor control

-> 

Replace motor control (Pos. 25)

11 Motor enable timeout

    Motor could not be enabled in time

-> 

Replace motor control (Pos. 25)

12 Invalid parameter motor error:

     Motor has received a invalid parameter

-> 

Replace motor control (Pos. 25)

-> 

Replace the hand held programming unit (Pos. 20)

13 Stop release error:

     The motor could not be released

-> 

Replace motor control (Pos. 25)

-> 

Replace the hand held programming unit (Pos. 20)

14 Unexpected motor Stop:

-> 

Check spiral cable and connector

-> 

Replace motor control (Pos. 25)

15 Motor disabled:

     Motor control disabled the motor.

-> 

Replace motor control (Pos. 25)

16 Wrong command in the motor :

-> 

Replace motor control (Pos. 25)

-> 

Replace the hand held programming unit (Pos. 20)

17 5V supply error:

     5V supply of motor control not sufficient

-> 

Replace motor control (Pos. 25)

18 Initialise error real time clock:

-> 

Replace the hand held programming unit (Pos. 20)

19 Communication error real time clock:

-> 

Replace the hand held programming unit (Pos. 20)

20 Error real time clock

:

-> 

Replace the hand held programming unit (Pos. 20)

3

21 Range exceeded:

The measured angle is out of the range of motion

-> 

Replace motor control (Pos. 25)

22 ROM error in the hand held programming unit:

     Memory error in the hand held programming unit

-> 

Replace the hand held programming unit (Pos. 20)

23 Invalid parameter:

     Internal error in the hand held programming unit

-> 

Replace the hand held programming unit (Pos. 20)

24 24V supply error motor control:

      Error in the 24V supply in the motor control

-> 

Replace motor control (Pos. 25)

-> 

Replace the power supply electronics (Pos. 28)

25 Bus error:

     Bus system error

-> 

Replace the spiral cable of the hand held

      programming unit

-> 

Replace the hand held programming unit (Pos. 20)

-> 

Replace motor control (Pos. 25)

26 24V supply hand held programming unit:

     24V supply of the hand held programming unit is
     defective

-> 

Replace the hand held programming unit (Pos. 20)

27 5V supply hand held programming unit:

     5V supply of the hand held programming unit is defective

-> 

Replace the hand held programming unit (Pos. 20)

28 3.3V supply hand held programming unit:

     3.3V supply of the hand held programming unit is
     defective

-> 

Replace the hand held programming unit (Pos. 20)

29 Calibration:

      The calibration data in the motor control are wrong.

-> 

Perform a calibration (see chapter 8.1)

30 Calibration error:

-> 

Repeat the calibration (see chapter 8.1)

-> 

Replace knee electronics (Pos. 5)

-> 

Replace motor control (Pos. 25)

31 Calibration timeout

:

-> 

Replace motor control (Pos. 25)

32 Motor enable error:

     The motor could not be enabled

-> 

Replace motor control (Pos. 25)

33 Motor disable error:

     The motor could not be disabled

-> 

Replace motor control (Pos. 25)

34 Motor stop error:

     Motor stop command timeout error

-> 

Replace motor control (Pos. 25)

35 Configuration error:

      Invalid configuration of the hand held programming unit

-> 

Replace the hand held programming unit (Pos. 20)

36 Parameter not valid chipcard

-> 

Replace the chipcard (Pos. 21)

Содержание ATRTOMOT-K1

Страница 1: ...Service Manual ARTROMOT K1...

Страница 2: ...for the hand held program ming unit have to be closed at all times When you assembling with electronic parts make sure to use ESD Electro Static Discharge equipment Only original chip cards may be us...

Страница 3: ...ld programming unit Pos 20 3 21 Range exceeded The measured angle is out of the range of motion Replace motor control Pos 25 22 ROM error in the hand held programming unit Memory error in the hand hel...

Страница 4: ...held programming unit Use the correct hand held programming unit Pos 20 46 Handset error internal communication Invalid interchip communication inside the hand held programming unit Replace the hand h...

Страница 5: ...iginal packaging for transport Put the hand held programming unit into the specified cutout of the styrofoam and fix it with adhesive tape Set the femur lenght on maximum and the lower leg setting on...

Страница 6: ...ts ARTROMOT K1 Display Control logic Motor driver Micro processor Motor Control Hand held programming unit Microprocessor Control logic Pot Control Motor Pot Power supply 24 V DV Optional card reader...

Страница 7: ...10 Belt loop 2 0013 087 11 Footrest complete with plate clamping lever belt 2 0037 032 Footrest 2 0013 300 Footplate 2 0037 168 Oval plate 2 0013 298 12 Fixation screw GN534 40 M6sw Rubber puffer comp...

Страница 8: ...trol 2 0037 901RevB 26 Motor 0 0037 010 27 Wire set 2 0037 005 28 Power supply electronics 0 0034 244 29 Lip 2 0037 105 Wire for lip 0 0031 300 30 Spindel complete 0 0037 203 Power cord EU version 0 0...

Страница 9: ...9 27 26 24 25 29 28 Position 30...

Страница 10: ...tion starts automatically The device will reach both maximum points and move between 10 bis 120 degrees with different speed Wait until the ARTROMOT K1 stops If the calibration was succesful the devic...

Страница 11: ...The symbol of the causer Hand held programming unit Motor Press or to see the entries of the error log Press STOP twice to leave the service menu General note to the error log Entries are always in en...

Страница 12: ...ion ARTROMOT K1 move in direction Extension Flexion ARTROMOT K1 move in direction Flexion until the ARTROMOT K1 reach 0 degrees Press START the calibration starts automatically The device will reach b...

Страница 13: ...and held programming unit Motor Press the control Extension last entry Flexion next entry to see the entries of the error log Press STOP twice to leave the service menu General note to the error log E...

Страница 14: ...make sure to use ESD Electro Static Discharge equipment Remove the housing cover see chapter 10 1 Pull out the connectors of the motor control Loosen the four screws Exchange the defective motor cont...

Страница 15: ...degrees Tolerance 5 degrees Check the angle in position 100 degrees Tolerance 5 degrees 3 Check the emergency off function Start the ARTROMOT K1 in any mode Press any key the ARTROMOT K1 will stop im...

Страница 16: ...ion 0 degrees Tolerance 5 degrees Check the angle in position 60 degrees Tolerance 5 degrees Check the angle in position 100 degrees Tolerance 5 degrees 3 Check the emergency off function Start the AR...

Страница 17: ...DIN EN 13485 ORMED Nr 018 829 01 ORMED GmbH Co KG Merzhauser Stra e 112 D 79100 Freiburg Tel 49 761 4566 01 Fax 49 0761 4566 5501 www ormed de E Mail info ormed de...

Отзывы: