10. How to perform repairs
10.1 How to remove the housing cover (Pos. 22 + 23).
Move ARTROMOT®-K1 in a position approximately
80 degrees.
Turn of the power OFF at the ARTROMOT®-K1 and
remove the power cord.
Move the ARTROMOT®-K1 in a stable side position.
Loosen the 4 outside torx screws to remove the hou-
sing cover K1 with sticker (Pos. 22).
Loosen the 2 torx screws to remove the housing cover
small (Pos. 23).
If you have to exchange any of the printed circuit boards
including the knee electronics, hand held programming
unit, motor control or power supply electronics you
have to perform a calibration.
10.2 How to exchange the motor control (Pos. 25).
ATTENTION!
When you assembling with electronic parts make
sure to use ESD (Electro Static Discharge) equipment.
Remove the housing cover, see chapter 10.1.
Pull out the connectors of the motor control.
Loosen the four screws.
Exchange the defective motor control and fix it with the
screws.
Put back in the connectors in the same position.
Rebuild the housing cover.
Perform a calibration.
Finally, a safety and function test has to be performed.
10.3 How to exchange the power supply electronics
(Pos. 28).
ATTENTION!
When you assembling with electronic parts make sure
to use ESD (Electro Static Discharge) equipment.
Remove the housing cover, see chapter 10.1.
Pull out the connectors of the power supply electronics.
Loosen the four screws.
Exchange the defective electronics and fix it with the
screws.
Put back in the connectors in the same position.
Rebuild the housing cover.
Perform a calibration.
Finally, a safety and function test has to be performed.