7
The ARTROMOT
®
-K4 CPM device
allows extension and flexion of the
knee joint in the range of -10-0-125
degrees and of the hip joint in the range
of 10-100 degrees.
The ARTROMOT
®
-K4 features a
hand-held programming unit that can
be used to program and store any
treatment values.
6
– The ARTROMOT
®
-K4 may only be
operated by authorised persons.
– Make sure that the patient is sup-
ported in an anatomically correct way.
Check the following settings /posi-
tioning:
1. Femur length
2. Knee joint axis
3. Hip joint axis
4. Calf length and leg rotation setting
5. Patient kits
– In case of patients who are adipose,
particular large or very small, you
should pay attention to the following:
– Avoid abrasion and pressure
– If necessary support the leg in a
slightly abductive position.
– The maximum continuous load on the
leg support element is 30 kg.
– Movement must always be free of
pain and irritation.
– The patient must be fully conscious
during instruction and when using the
splint.
– The doctor or therapist must decide
on a case-to-case basis whether the
device can be used with the patient.
– The hand-held programming unit
should be explained to the patient
and must be located within the
patient’s reach, so that the therapy
can be interrupted if necessary.
– Make sure that the characteristic
values of your power supply corre-
spond to the voltage and frequency
data indicated on the ID plate.
– Only connect the ARTROMOT
®
-K4
to correctly installed safty sockets.
– Repair and maintenance work may
only be carried out by authorised
persons, as otherwise all warranty
services and liabilities shall be void.
– Perform regular checks on all compo-
nents for possible damage or loose
connections.
– Damaged or worn parts should be
replaced immediately with original
spare parts by an authorised spe-
cialist.
– Before cleaning and repair disconnect
the device from the main socket.
– When carrying out any work on the
device, never allow liquids to get
inside the housing or the hand-held
programming unit.
– Only use the AC-AC adapter supplied
with the unit.
To disconnect the device from mains,
unplug the AC-AC adapter from the wall
socket.
Note: Fold out page 2
1
Button for heigth adjustment of hip
pivot point
2
Spare tube
3
Release tubes for square tube
4
Patient kit strapes
5
Coiled cord
6
Controller cable
7
Hand-held programming unit
8
Knob for length adjustment
of lower leg
9
Base
10 Knob for angle adjustment of foot
inclination
11 Knob for rotation footplate
12 Power adapter
13 Cable of power adapter
14 Socket for power adapter
15 Main switch
16 Footplate with patient kit
17 Lower leg support
18 Lower leg patient kit
19 Knee pivot point
20 Thigh patient kit
21 Thigh support
22 Knob for femur length adjustment
23 Hip axis pivot point
Explanation of symbols
Alternating current
Protective system Type B
Power off
Power on
Device off
Device on
2. Description of the ARTROMOT
®
-K4
Explanation of the functioning
elements
3. Safety instructions
PRECAUTION!
these instructions must be read
before start-up!
PRECAUTION!
Before treatment begins, a test
run involving several movement
cycles should be carried out first
without and then with the patient.
PRECAUTION!
The ARTROMOT
®
-K4 may only be
operated with the attached power
supply NTEV20.