7

The ARTROMOT

®

-K4 CPM device

allows extension and flexion of the 
knee joint in the range of -10-0-125
degrees and of the hip joint in the range
of 10-100 degrees.

The ARTROMOT

®

-K4 features a 

hand-held programming unit that can
be used to program and store any 
treatment values.

6

– The ARTROMOT

®

-K4 may only be

operated by authorised persons.

– Make sure that the patient is sup-

ported in an anatomically correct way.
Check the following settings /posi-
tioning:

1. Femur length

2. Knee joint axis

3. Hip joint axis

4. Calf length and leg rotation setting

5. Patient kits

– In case of patients who are adipose,

particular large or very small, you
should pay attention to the following:

– Avoid abrasion and pressure

– If necessary support the leg in a

slightly abductive position.

– The maximum continuous load on the

leg support element is 30 kg.

– Movement must always be free of

pain and irritation.

– The patient must be fully conscious

during instruction and when using the
splint.

– The doctor or therapist must decide

on a case-to-case basis whether the
device can be used with the patient.

– The hand-held programming unit

should be explained to the patient 
and must be located within the
patient’s reach
, so that the therapy
can be interrupted if necessary.

– Make sure that the characteristic 

values of your power supply corre-
spond to the voltage and frequency
data indicated on the ID plate.

– Only connect the ARTROMOT

®

-K4

to correctly installed safty sockets.

– Repair and maintenance work may

only be carried out by authorised 
persons, as otherwise all warranty
services and liabilities shall be void.

– Perform regular checks on all compo-

nents for possible damage or loose
connections.

– Damaged or worn parts should be

replaced immediately with original
spare parts by an authorised spe-
cialist.

– Before cleaning and repair disconnect

the device from the main socket.

– When carrying out any work on the

device, never allow liquids to get
inside the housing or the hand-held
programming unit.

– Only use the AC-AC adapter supplied

with the unit.

To disconnect the device from mains,
unplug the AC-AC adapter from the wall
socket.

Note: Fold out page 2

1

Button for heigth adjustment of hip
pivot point

2

Spare tube

3

Release tubes for square tube

4

Patient kit strapes

5

Coiled cord

6

Controller cable

7

Hand-held programming unit

8

Knob for length adjustment 
of lower leg

9

Base

10 Knob for angle adjustment of foot

inclination

11 Knob for rotation footplate

12 Power adapter

13 Cable of power adapter

14 Socket for power adapter

15 Main switch

16 Footplate with patient kit

17 Lower leg support

18 Lower leg patient kit

19 Knee pivot point

20 Thigh patient kit

21 Thigh support

22 Knob for femur length adjustment

23 Hip axis pivot point

Explanation of symbols

Alternating current

Protective system Type B

Power off

Power on

Device off

Device on

2. Description of the ARTROMOT

®

-K4

Explanation of the functioning 
elements 

3. Safety instructions

PRECAUTION!

these instructions must be read
before start-up!

PRECAUTION!

Before treatment begins, a test 
run involving several movement
cycles should be carried out first
without and then with the patient.

PRECAUTION!

The ARTROMOT

®

-K4 may only be

operated with the attached power
supply NTEV20.

Содержание ARTROMOT-K4

Страница 1: ...Operation Manual ARTROMOT K4...

Страница 2: ...2 12 13 17 11 10 9 14 8 1 2 5 15 16 3 4 7 6 18 19 19 20 21 22 23 3 Device description Figures Fig 2 Fig 3 Fig 4 Fig 5 Fig 6 Fig 7 Fig 8...

Страница 3: ...hroscopy procedures in combination with synovectomy arthrolysis or other intra articular measures mobilsations of joints in narcosis operative treatments on fractures pseudarthrosis and inversion oper...

Страница 4: ...persons as otherwise all warranty services and liabilities shall be void Perform regular checks on all compo nents for possible damage or loose connections Damaged or worn parts should be replaced im...

Страница 5: ...OMOT K4 features a true anatomical knee and hip axis for maximum patient comfort ARTROMOT K4 has to be set up either for the right or left leg The device can be converted quickly The procedure is easi...

Страница 6: ...Speed Minimum setting for speed 1 Maximum setting for speed 100 Special functions are Center warm up Full speed motion double speed setting Runtime patient runtime Device runtime Programming the spec...

Страница 7: ...lative humidity 20 to 85 Air pressure 700 hPa to 1060 hPa Operational Conditions Surrounding temperature 50O F to 104O F Relative humidity 30 to 75 Air pressure 700 hPa to 1060 hPa 13 12 6 Maintenance...

Страница 8: ...dical Devices Part 1 and 2 additional norm electromagnetic compatibility requirements and testing The adherence to the standard specifications entitles to marking of these devices with CE 0297 0297 Fr...

Отзывы: