background image

9

8

– Fix patient kit (18) for the lower leg

and patient kit (20) for the upper leg
by using the velcro tapes (figure 6
and figure 7).

– Control correct adjustment. Exercise

only in painfree range of motion.
Patient should be positioned with
maximum comfort.

Note: Fold out pages 2 and 3. 
To get a better understanding of the
individual steps.

– Connect the power adapter (7) to a

to a safety socket (120 Volt, 60 Hertz) 

– Turn on the device with the main

switch (15)

Set the device at knee-angle position
that is not likely to cause the patient
any pain.

Positioning the upper leg

– Open the black knurled knob (22),

push button and move thigh support
(21) to the desired length (figure 2).

– For correct alignment of the hip axis

pivot point (23) and anatomical hip
axis of the patient pull button (1) and
align pivot point (23) to height of
trochanter major of the patient.

Positioning the lower leg

– Loosen the two knobs (8), move the

foot support horizontally and adjust
precisely to the patients lower leg
length (figure 3).

Positioning of foot dorsi-/ plantar
flexion

– Loosen the two knobs (10) and adjust

the foot plate at a comfortable angle
(figure 4).

Positioning of foot rotation

– Loosen the knurled knob (11) and

move the foot plate into the required
rotation position (figure 5).

ARTROMOT

®

-K4 features a true

anatomical knee and hip axis for 
maximum patient comfort.

ARTROMOT

®

-K4 has to be set up

either for the right or left leg.

The device can be converted quickly.
The procedure is easiest at an angle 
of approximatly 80 – 90 degrees
(section 5.1.1).

– Hand-held programming unit (7) is in

STOP mode.

– Pull button for height adjustment

mechanism (1) and remove thigh 
support (21) (figure 9).

– Hold the thigh support. Release

length adjustment mechanism 
(1/4 rotation) from the bayonet lock
(figure 10).
Remove entire part and slide into 
the opposite side and fix in place with 
the bayonet lock.

– Press buttons (3) simultaneously and

pull square tube (2) associated with
the heigth adjustment mechanism out
from profile (figure 11). Slide into the
opposite profile until it “clicks” home
audibly.

– Slide the height adjustment elements

together again and allow hinge to
click home at the same height as the
turning point of the hip (figure 12).

Fig. 12

Fig. 9

4.1 Connecting 

the device

4.2 Adjusting the 

femur length

4. Adjusting the device

4.3 Adjusting the 

patient kit

PRECAUTION!

The knee and hip axis of the
ARTROMOT

®

-K4 should align with

the patients knee and hip axis 
(figure 8).
After adjustments have been
made, perform several test runs.
When correctly adjusted, there
should be no excursion of the 
knee and hip joint during motion.

4.4 Conversion

Fig. 10

Fig. 11

PRECAUTION!

For correct insert and lock posi-
tion of the bayonet lock refer to
sticker on the device.

Содержание ARTROMOT-K4

Страница 1: ...Operation Manual ARTROMOT K4...

Страница 2: ...2 12 13 17 11 10 9 14 8 1 2 5 15 16 3 4 7 6 18 19 19 20 21 22 23 3 Device description Figures Fig 2 Fig 3 Fig 4 Fig 5 Fig 6 Fig 7 Fig 8...

Страница 3: ...hroscopy procedures in combination with synovectomy arthrolysis or other intra articular measures mobilsations of joints in narcosis operative treatments on fractures pseudarthrosis and inversion oper...

Страница 4: ...persons as otherwise all warranty services and liabilities shall be void Perform regular checks on all compo nents for possible damage or loose connections Damaged or worn parts should be replaced im...

Страница 5: ...OMOT K4 features a true anatomical knee and hip axis for maximum patient comfort ARTROMOT K4 has to be set up either for the right or left leg The device can be converted quickly The procedure is easi...

Страница 6: ...Speed Minimum setting for speed 1 Maximum setting for speed 100 Special functions are Center warm up Full speed motion double speed setting Runtime patient runtime Device runtime Programming the spec...

Страница 7: ...lative humidity 20 to 85 Air pressure 700 hPa to 1060 hPa Operational Conditions Surrounding temperature 50O F to 104O F Relative humidity 30 to 75 Air pressure 700 hPa to 1060 hPa 13 12 6 Maintenance...

Страница 8: ...dical Devices Part 1 and 2 additional norm electromagnetic compatibility requirements and testing The adherence to the standard specifications entitles to marking of these devices with CE 0297 0297 Fr...

Отзывы: