14
9.1
Return
In order to avoid damage due to transport and shipping in the event of a return, please use only the
original shipping packaging.
For shipping and transport, always connect the pressure tester to the tester connection with the valve
open!
10 Disposal
Environmentally friendly disposal according to EU Directive 2012/19/EU. The appliance
contains electronic components. To prevent environmental risks or hazards due to im-
proper disposal, the product, including accessories, must be disposed of in accordance
with the applicable EU directives 2012/19/EU. Disposal can be carried out by the manu-
facturer.
For this purpose, please send to the manufacturer at:
Orlvision GmbH, Gewerbestraße 17, D-35633 Lahnau.
Disposal in household waste is prohibited.
11 Electromagnetic compatibility
11.1
Information on the operating environment:
The Video Rhino Laryngoscope is intended for low-interference RF environments such as doctors' of-
fices. No shielded location is required.
11.2
Information on the performance characteristics
Key features of
the Video Rhino Laryngoscope are: Display of images of the examination re-
gion (the nose, pharynx and larynx, nasal cavities and nasopharynx). If there is strong electro-
magnetic interference, the image quality may be impaired.
Warning:
Use of this unit immediately adjacent to other units or with other units stacked to-
gether should be avoided as this may cause interference.
Cables, transformers and accessories that can be replaced without affecting the EMC: None
Warning: The
use of other accessories (especially PC) may lead to incorrect operation
Warning
: Portable communication (radio) equipment operating in close proximity may result in
incorrect operation
11.3
Electrical immunity (immunity)
Parameter
Data RSX-USB
Data RSX-P
Static electricity discharge (ESD)
according to EN 61000-4-2, level
according to EN 60601-1-2
+/- 8 kV Contact discharge
+/- 15 kV Air discharge
11.4
Immunity test
Parameter
Data RSX-USB
Data RSX-P
Immunity to electromagnetic radiation
according to EN 61000-4-3, level ac-
cording to EN 60601-1-2
f = 80 MHz to 2.7 GHz
10V/M; 3 V/m
According to EN 60601-1-2 chapter
8.10; / PM, 18 Hz or 217 Hz
380 MHz - 5.8 GHz
9 - 28 V/m