12
Automatic cleaning and disinfection process with programme no. 22 NORMAL-PAA with the following
settings:
Step Pre-cleaning for 4 minutes
Step Emptying
Step Cleaning 0.5% at 37°C for 6 minutes
Step Cleaning 0.5% at 43°C for 6 minutes
Step Emptying
Step Intermediate rinse for 2 minutes
Step chemical disinfection with 1% disinfectant at 25°C for 10 minutes.
Step Emptying
Step Final cleaning at 20°C for 4 minutes
Attention:
Usually the laryngoscope is connected to the automatic pressure monitoring during machine cleaning.
If this is not done, we recommend using the optionally available cover cap M-860-0003-0086-P to pro-
tect the valve connection.
It is also recommended to use this cover cap for manual cleaning.
Fig. 5, Cover cap Article number M-860-0003-0086-P
Figure 6: Valve cover cap fitted
8.3
Leak test
Attention:
The leak test must be carried out before
each reprocessing
!
Caution: The connection tube of the leak tester and the tester connection on the laryngoscope must
be dry!