background image

68805-M

ARC LAMP POWER SUPPLY

1

I.

SAFETY

SAFETY NOTES:

- Connections to or from any electrical device must never be made while the device is switched
on.

- The Oriel Model 68805 Power Supply is such a device. It has an open circuit voltage of >160
volts when the start switch is pressed, and could be dangerous if care is not exercised when the
power supply is switched on. Please follow these precautions when operating the power supply:

Make all input and output connections before turning the power supply on.

Do not use the power supply without its cover.

Ensure that the output terminals are properly connected to the lamp housing or
lamp socket before turning on the power supply.

Don't open the power supply or disconnect the output leads while the power
supply is on.

- Follow the lamp or lamp house manufacturer's recommendations and precautions when handling
and using lamps.

- Never touch the lamp's quartz envelope with uncovered fingers; definitely not when operating
the lamp or immediately after shutdown. Also, follow this precaution during lamp installation to
avoid damage to the lamp.

- Do not start the lamp with the output control set above the current rating for the lamp.

- The lamp is a source of constant current. To prevent overpowering of the lamp as it ages,
monitor its current on regular basis.

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Содержание 68805

Страница 1: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Страница 2: ...Avenue De Laponie Im Tiefen See 58 1 Mole Business Park Z A De Courtaboeuf D 64293 Darmstadt Leatherhead 91951 Les Ulis Cedex Germany Surrey KT22 7AU France England Phone 1 69 07 20 20 Phone 06151 880...

Страница 3: ...DUCTION 2 III SPECIFICATIONS 4 IV CONTROLS AND CONNECTIONS 5 V CIRCUIT DESCRIPTION 11 VI CALIBRATION AND TROUBLESHOOTING 14 WARRANTY AND RETURNS Artisan Technology Group Quality Instrumentation Guaran...

Страница 4: ...connected to the lamp housing or lamp socket before turning on the power supply Don t open the power supply or disconnect the output leads while the power supply is on Follow the lamp or lamp house m...

Страница 5: ...t current power supply regulates current even into a short circuit and the model 68805 is safely limited to 12 amperes or 300 watts Also the inrush current to the lamp is limited to a maximum of 12 am...

Страница 6: ...will allow it to operate all arc lamps that are rated for DC operation in this voltage current and power range The main power supply section uses offline switching technology with the resultant increa...

Страница 7: ...ition 150 volts unloaded Operating Load dependent Total output power should not exceed 240 watts Light Output Ripple Resistive Load 0 25 R M S 40 Hz to 40 kHz Arc Lamp Load Dependent on lamp and opera...

Страница 8: ...lamps through improper operation FRONT PANEL Figure 3 1 shows the layout for the front panel METER AND VOLTAGE WATTS AND AMPS SWITCH The analog meter is scaled to display the output voltage current an...

Страница 9: ...lay the value of the output current that will be delivered to the lamp after lamp start POWER SWITCH This is the AC mains power switch In the ON position ac power will be switched into the main circui...

Страница 10: ...MAINS INPUT J304 J304 is an IEC style socket which will accept power cords for worldwide use of the power supply The AC mains cord is connected here LINE VOLTAGE SELECTOR S301 S301 sets the input vol...

Страница 11: ...HF screw on connectors are the connections for the main DC output to the load J306 is the positive and J307 is the negative The low voltage side of an external ignitor or a lamp housing with built in...

Страница 12: ...to the output current This signal should be buffered with a unity gain amplifier if it is going to be used Pin 3 Input control and reference output common Pin 4 Not used Pin 5 Remote start input A con...

Страница 13: ...control knob is set to the full counter clockwise position Turn on the power supply Place the meter select switch to VOLTS and observe that the output is 100 volts DC Release the meter switch Press t...

Страница 14: ...contact closure A complete connection at J301 will energize K1 and allow mains power to be applied to the main power supply module 27 20 007 The power supply module will then begin to operate and con...

Страница 15: ...control Meter Board 68805 1210 Please refer to schematic number 68805 3 1211 for this circuitry J1 pins 7 and 9 provide the AC for this circuit This is rectified and filtered by CR1 C1 and C3 The raw...

Страница 16: ...8805 3 1211 for their location Ignitor Drive 68810 3 1601 Please refer to schematic number 68810 3 1601 for this circuitry Ignitor Drive and Output Monitor Circuit The ignitor drive PCB contains circu...

Страница 17: ...hat the output current as viewed on the external ammeter is an even value Hold the meter select switch in the AMPS position and trim R6 so that the front panel meter display agrees with the external a...

Страница 18: ...ponents which are readily available No attempt should be made to repair the power supply module to the component level Simply replace and return the defective module to Oriel A flow chart of a suggest...

Страница 19: ...r board is OK Shut down connect the resisitive load and turn the unit on Measure 5V at J102 pin 1 Replace module if not present Measure 0 5V range at J102 pin 2 Replace module if not present Does the...

Страница 20: ...er ONLY IF the defective goods are returned freight prepaid to an Oriel Instruments designated facility If goods are not returned to Oriel Instruments and user chooses to have repairs made at their pr...

Страница 21: ...and warranted use and performance of the goods recipient of the goods accepts full responsibility for restoring the goods to their condition upon original delivery and for assuming all costs and charg...

Страница 22: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...