| INSTRUCTIONS FOR USE
OM 402PWR
9.
ERROR STATEMENTS
ERROR
CAUSE
ELIMINATION
E. DIS._
Number is too small (large negative) to be displayed
change DP setting, channel constant setting
E. DIS.-
Number is too large to be displayed
change DP setting, channel constant setting
E. TAB._
Number is outside the table range
increase table values, change input setting
(channel constant setting)
E. TAB.-
Number is outside the table range
increase table values, change input setting
(channel constant setting)
E. INP. _
Input quantity is smaller than permitted input quan-
tity range
change input signal value or input (range) setting
E. INP.-
Input quantity is larger than permitted input quantity
range
change input signal value or input (range) setting
E. HW.
A part of the instrument does not work properly
send the instrument for repair
E. EE.
Data in EEPROM corrupted
perform restoration of manufacture setting, upon repeated
error statement send instrument for repair
E.. SET.
Data in EEPROM outside the range
perform restoration of manufacture setting, upon repeated
error statement send instrument for repair
E. CLR
Memory was empty (presetting carried out)
upon repeated error statement send instrument for repair,
possible failure in calibration
E.. OUT.
Analogue output current loop disconnected
check wire connection
Содержание OM 402PWR
Страница 9: ...INSTRUCTIONS FOR USE OM 402PWR 9 3 INSTRUMENT CONECTION ...
Страница 31: ...INSTRUCTIONS FOR USE OM 402PWR 31 5 SETTING LIGHT ...
Страница 45: ...INSTRUCTIONS FOR USE OM 402PWR 45 6 SETTING PROFI ...
Страница 71: ...INSTRUCTIONS FOR USE OM 402PWR 71 6 SETTING PROFI ...
Страница 82: ...82 INSTRUCTIONS FOR USE OM 402PWR NOTE ...
Страница 83: ...INSTRUCTIONS FOR USE OM 402PWR 83 NOTE NOTE ...