INSTRUCTIONS FOR USE
OM 402LC
|
23
C. MIN
Calibration of the
beginning of the
measuring range
- prior performing any changes you will be
asked to confirm your selection “YES”
- prior confirmation of the selection the
reference signal has to be connected
C. MAX
Calibration of the end of
the measuring range
- prior performing any changes you will be
asked to confirm your selection “YES”
- prior confirmation of the selection the
reference signal has to be connected
Calibration of the end of the range > C. MAX
Example
Calibration of the beginning of the range > C. MIN
Example
C. MAX
YES
flashing legend
C. MIN
YES
flashing legend
YES
C. MAX
YES
LANG.
5
light
light
SETTING
After incorrect client calibration it is always possible to restore
manufacture calibration (“SERVIC./RESTOR/CALIB.”)
!
Manual calibration:
MAX Sensor
range
SENSE Sensor
sensitiveness
Automatic calibration
(after calibartion in menu “SERVIC./CALIB.”):
MIN
Size of load with which minimum
calibration was performed
MAX
Size of load with which maximum
calibration was performed
- upon maximum calibration we recom-
mend the reference load value in the upper third of
the measuring range
!