INSTRUCTIONS FOR USE
OM 402LC
|
55
6.4.2
Restoration of manufacture setting
Jobs performed
Restore
Calibration
Setting
cancels USER menu rights
deletes table of items order in USER - LIGHT menu
adds items from manufcture to LIGHT menu
deletes data stored in FLASH
cancels or linearization tables
clears tare
clears conduct resistances
restore manufacture calibration
restore manufacture setting
CHANNE.
SERVIC.
OUTPUT.
MENU
RESTOR.
LANG.
INPUTS
N. PASS.
YES
CALIB.
IDENT.
RE. CAL.
RE. SET.
SAVE
RESTOR.
Restoration of
manufacture setting
- in the event of error setting or calibration,
manufacture setting may be restored.
RE. CAL. .
Restoration of
manufacture calibration
of the instrument
- prior executing the changes you will be
asked to confirm you selection
„YES“
RE. SET.
Restoration of instrument
manufacture setting
FIRM.
Restoration of instrument
manufacture setting
- generating the manufacture setting
(items marked DEF)
USER
Restoration of instrument
user setting
- generating the instrument user setting, i.e.
setting stored under SERVIC./RESTOR./
SAVE
SAVE
Save instrument user
setting
- storing the user setting allows the operator
to restore it in future if needed
CHANNE.
SERVIC.
OUTPUT.
MENU
RESTOR.
LANG.
INPUTS
N. PASS.
FIRM.
CALIB.
IDENT.
RE. CAL.
RE. SET.
SAVE
USER
After restoration the instrument switches off for
couple seconds
!
profi
profi
6
SETTING