Technical Data
13
-
2.2 Factory default
The CT G5L pyrometer is delivered with the following factory settings:
Signal output object temperature
0-10 V
Emissivity
0,920
Transmissivity
1,000
Averaging (AVG)
0,2 s
Smart Averaging
inactive
Peak hold
active (Hold time: 15 s)
Valley hold
Inactive
Lower limit temperature range [°C]
100
Upper limit temperature range [°C]
900
Lower alarm limit [°C] (normally closed)
200
Upper alarm limit [°C] (normally open)
500
Lower limit temperature range [°C]
0 V
Upper limit temperature range [°C]
10 V
Temperature unit
°C
Ambient temperature compensation
internal head temperature probe
Содержание Top-down GIS