40
Min. interval
controls that the flag will not be released before given time period even thought it might be
required by the imager. If the
Max. interval
is defined the flag will be always released after certain time period
even if not required by the imager.
Emissivity / Ambient temperature
You may enter fixed emissivity/ambient temperature value.
By default the ambient temperature value is delivered by the camera’s internal sensor
In case that the values are controlled through Process Interface (PIF) the here defined values will be ignored.
Reference temperature
There are two options available to define the reference temperature – by entering fixed value or using the
Process
Interface input (PIF in)
.
Using comparisons of reference temperature source automatic or user-defined adjustments of object
temperature can be carried out
Temperature range
Temperature range from 20°C to 100°C, from 0°C to 250°C or from150°C bis 900°C are available.