Mechanical Installation
25
-
3.2 Laser Sighting
The integrated laser sighting supports the alignment and focusing of the optics. The
size of the laser dot is equal to the real measurement spot size at any distance.
The laser can be activated/ deactivated via the programming keys on the unit or via
the software. If the laser is activated a yellow LED will shine (beside temperature
display).
[►4 Electrical Installation, 6 Operating]
At ambient temperatures >50 °C inside the electronic box the laser will switch
off automatically.
Do not point the laser directly at the eyes of persons or animals! Do not stare into the
laser beam. Avoid indirect exposure via reflective surfaces!
The laser is not suitable for continuous operation.
Before you open the box, remove the fiber optical cable from the electronics or disassemble the
optical head the sensor must be switched off and disconnected from the power supply and USB
interface!
Содержание CTratio 1M Series
Страница 1: ...Operator s Manual optris CTratio 1M 2M Fiber Optics Ratio Thermometer...
Страница 54: ...54 Appendix D Declaration of Conformity...
Страница 55: ......
Страница 56: ...optris CTratio MA E2019 12 A...