Installation
41
-
Figure 23:
Terminal block
Designation
1. Ground
2. Shutter Status LED
3. Not assigned
4. Analog Input for the shutters
– remote control
5. Analog Input for the shutters
– external trigger
6. to 9: 24 V DC Power supply
10. to 13: Ground
The wiring of the furnace signal on the control cabinet is done under connection 5 (Input) and connection 10
(ground).
The input voltage range is 0 / max. 24 V. An open input is interpreted as a high signal. The signal
must be switched to ground (0 V). If your system provides an open input as a low signal, the
alarm mode must be changed from normally open to normally closed on the CTL 4M via
CompactPlusConnect and the system must be restarted.