36
-
Figure 18: Layout BUGIS_Operation
Now start the layout BUGIS_Operation. The
shutters must be open for this step. To do
this, go to the Line scanner configuration via
Tools - Line scanner mode - Line scanner
settings
under the tab
General
click on the
Continuous
mode. After the fine adjustment
set the Line scanner configuration mode back
to
External triggered.
Now a few settings have to be made. The
scan line must be positioned correctly. The
scan line can be moved in the lower right
window (line scanner live view) in the
specified 5 points by clicking and dropping.
Alternatively, coordinates can also be entered via the line scan camera configuration via Line scanner
configuration.
Make absolutely sure that the width of the scan line is at least as wide as the glass passes through. This
ensures that the entire glass is scanned.