background image

 

 

V 1.1 - eng 

 

 

3.

 

Precisely adjust the mount position by means of adding/removing back focal adjustment 

spacers until you find the best focusing position. 

4.

 

Adjust the phase angle. Once you find the right phase, lock the screws (M2 metric thread). 

5.3.

 

How to change the magnification with MTDV motion controller 

Be sure to correctly load the lens XML configuration file (see sections 5.1.1 and 5.1.3

 

for details). 

Now the four different magnifications of the TCZRS zoom lens are saved into the memory slots of 

the MTDV motion controller. 

Go 

in the “Operation” web page of the MTDV GUI  (see  Figure  3) at  the “Memory Save/Recall” 

section. Type a number in the range of memory slots you choose when importing XML and click 

“Load” button to change the magnification.  

6.

 

CENTER POSITION REPEATABILITY 

Center position repeatability is defined as the positional stability of a reference point at the same 

magnification, keeping the same measurement configuration. 

Choosing a reference point C

1

  (X

1

;Y

1

), after n measurements at the same  magnification, the 

distribution of reference point positions (C

1

,C

2

,…, C

n

) has a mean C

m

(X

m

;Y

m

) that is the Center of the 

repeatability area. The radius  of this area is the maximum distance between the mean of the 

distribution and any other reference point registered in testing. 

7.

 

CENTER DISPLACEMENT 

Every zoom system shows different image centering at different magnification settings. 

Choosing a reference point at a given magnification, when magnification is  changed this point 

shifts.  The maximum distance between reference points  at different magnifications is  the  center 

displacement. To better understand this effect, you can refer to the following picture.  

To eliminate this effect, you can select a reference point at the image center when the lens is set at 

its maximum magnification. Then you can measure the x and y image center displacements at the 

other three magnifications. 

The x and y (dx and dy) displacements must be recorded together with other  calibration 

parameters specific for each zoom configuration (magnification factor,  residual distortion 

correction parameters...) 

The maximum x or y displacement will define the width and height of a crop-window which will 

show the image center always perfectly aligned. 

 

 

Содержание TCZR036S

Страница 1: ...V 1 1 eng 8x Bi Telecentric zoom lenses with motorized control TCZR036S TCZR072S OPTICS INSTRUCTION MANUAL...

Страница 2: ...CIFICATIONS 4 4 1 Specifications table 4 4 2 Connector 5 5 INSTRUCTIONS FOR USE WITH MTDV MOTION CONTROLLER 6 5 1 First use of TCZRS with MTDV motion controller 6 5 2 Clamping turning on and focusing...

Страница 3: ...ver might be permanently damaged if deployed in the presence of strong vibrations and impulsive forces TCZRS must be properly clamped before usage Zoom Revolver must be mechanically constrained by ade...

Страница 4: ...04 0 8 40 2 000 11 0 08 0 2 35 TCZR072S 0 125 11 157 8 16 0 05 0 1 53 0 35 C Yes 279 7 144 103 2850 0 250 11 0 08 13 2 40 0 500 11 0 05 3 3 40 1 000 11 0 07 0 8 35 Notes 1 Working F the real F of a l...

Страница 5: ...ibration sources and impulse forces 4 2 Connector Figure 1 DB15 HD connector male panel view PIN NAME DESCRIPTION 1 5V 5V encoder power supply 2 GND 0V encoder reference ground 3 ENC_A Encoder quadrat...

Страница 6: ...file Download the XML configuration file of your zoom lens on the web page https www opto e com configuration_files by typing the serial number of the lens and the Opto Security Number and by clickin...

Страница 7: ...set to save this information For example if you choose 0 motor positions are saved into 0 1 2 and 3 memory slots If you choose 21 motor positions are saved into 21 22 23 and 24 memory slots Click Impo...

Страница 8: ...rms a homing procedure and it stops to the last reached position After clamping attaching the camera and turning on the device please follow step by step the focusing procedure explained below 1 Adjus...

Страница 9: ...fication the distribution of reference point positions C1 C2 Cn has a mean Cm Xm Ym that is the Center of the repeatability area The radius of this area is the maximum distance between the mean of the...

Страница 10: ...V 1 1 eng 10...

Отзывы: