                                                                                                    A23.0910     

 

www.optoedu.com

  

Page 6 of 14

 

[email protected]

 

 

2-2

 

Assembling Steps 

 
 
 
 
 
 
 
 
 
 
 
 

 

Fig. 2 

 
 
 

 

 

Fig. 3 

 

 

Fig. 4 

 

2-2-1

 

Assemble the Focusing Bracket Group 

Loosen  the  lock  knob

① 

of  the  focusing  bracket  group, 

insert  the  hole

② 

of  focusing  bracket  group  into  the 

column

,  until  the  focusing  bracket     group  to   the 

lowest position. Then tighten the lock knob

① 

to prevent 

the    microscope from turning. 

Slowly insert it, do not force too much. 

◎ 

To  avoid  the  microscope  turnover,  the  focusing 

bracket  group  must  assemble  on  the  same  side  to  the 

worktable  of  base  group,  as  the  left  view  of  Fig.  3.  If 

assemble the focusing  bracket group on the wrong side, 

the microscope will turn over as the right view of Fig. 3. 

 
 
 

2-2-2

 

Assemble the Microscope Body 

1)

 

Remove  the  decorative  cover

① 

of  the  focus  bracket 

group,  and  loosen  the  M4  inner  hexagon  screw  with  a 

M4 inner hexagon spanner. (See Fig. 4) 

2)

 

Match  the  dovetail  interface

③ 

of  the   microscope 

body  with  the  dovetail  groove

② 

of  the  focus  bracket 

group,  and  insert  it  from  top  to  bottom  as  the  direction 

shown in the figure. 

3)

 

Tighten  the  M4  inner  hexagon  screw  on  the  focus 

bracket  group  with  a  M4  inner  hexagon  spanner,    then 

put the decorative cover

① 

back. 

 
 

 

2-2-3

 

Assemble the Objective 

1) Rotate the objective as the arrowhead direction shown 
in  the  figure,  and  screw  the  objective 

② 

into  the 

objective installation screw hole

. (See Fig. 5) 

 

Fig. 5 

Содержание A23.0910

Страница 1: ...cnoec com To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly before start to use your instrument A23 0910 Zoom Stereo Microscope Parallel Light 0...

Страница 2: ...djust the Focusing Tension 9 3 4 Adjust the Diopter and Focus 10 3 5 Adjust the Interpupillary Distance 10 3 6 Adjust the Elevation Angle of Observation Head 10 3 7 Turn on off the Magnification Fixed...

Страница 3: ...l power adapter is used Wide voltage range is supported as 100 240V AC 5 Use the special wire supplied by our company 2 Maintenance 1 Keep all the surface of the lens clean Wipe the lens gently with a...

Страница 4: ...Pole Aperture Diaphragm Adjusting Rod Tension Adjusting Knob Main Objective Worktable 1 Components A23 0 Microscope Body Focusing Bracket Group Coarse Focusing Knob Clip Base Group Trinocular Head Ey...

Страница 5: ...embling Scheme to describe how to assemble the components and the numbers denote the assembling order Before assembling make sure there is no dust dirt or other materials which will disturb it Assembl...

Страница 6: ...the focusing bracket group on the wrong side the microscope will turn over as the right view of Fig 3 2 2 2 Assemble the Microscope Body 1 Remove the decorative cover of the focus bracket group and l...

Страница 7: ...hes the bottom Lock the eyepiece by the lock screw See Fig 7 Fig 7 Fig 8 Install or Remove the Reticle 1 Unscrew the pressure ring from eyepiece see Fig 8 2 Clean the reticle then install it into the...

Страница 8: ...the optical fiber into the inner hole of the power supply box to the bottom and align the flat surface with the lock screw then lock it with the lock screw 3 Insert the other end of the optical fiber...

Страница 9: ...stronger 1 Adjusting the adjustment knob of the reflector See Fig 14 can change the reflector surface also can change the lighting range after reflected 2 The adjustment knob of the reflector can be m...

Страница 10: ...knob to the maximum magnification if the image is not clear enough when observed with the right eyepiece repeat Step 3 5 as above 6 Then rotate the zoom knob to the minimum magnification turn the diop...

Страница 11: ...to the end with an inner hexagon spanner See Fig 19 Every magnification value on the zoom knob has a ticking sound fixed position 2 To turn off the ticking sound magnification fixed positon fuction r...

Страница 12: ...ead which can used for binocular observation See Fig 22 When the light path selecting pole is pulled to the outmost all the light will enter the trinocular head which can used for TV Photography Fig 2...

Страница 13: ...up onto the base with a M3 sunk key make the flat side of polarizer group close to the pillar then put back the worktable See Fig 24 2 Loop the analyzer group on the outside diameter of main objective...

Страница 14: ...ght field of view doesn t superpose with the left one Interpupillary distance is incorrect Adjust the interpupillary distance Diopter adjustment is incorrect Adjust the diopter The eyepiece for the ri...

Отзывы: