Maintenance 

5.1 

Clean:

 

(1)  Don’t touch the lens with hand, Dust on lens should be cleaned by soft brush or 

absorbent cotton or cleaned by absorbent cotton, lens paper with the mixture of 
alcohol and ether (proportion 1:4). 

(2)  Alcohol and ether all are burnt easily, please take them away from fire. Be careful for 

turn on and off power. 

(3)  Don’t clean painted metal and galvanizing metal with organic solvent such as 

alcohol, ether or the mixture of the both. Silicon cloth or soft cleaning preparation is 
suggested to clean it. 

(4)  Plastic should be cleaned by soft cloth with clear water. 

 
5.2 

Application environment 

(1)  Should be in cold and dry environment , without shaking ,dust and steam without 

acid-base and other corrosive gas

 

(2)  Normal operation condition

indoor temperature

0

℃~

40

 

maximum 

 

relative 

humidity

85% 

(3)  High humidity area

recommend

 

  with dehydrating unit in case of mildew and frog. 

 
5.3 

Bulb Replacement

 

A

 . 

Replacement of halogen.

 

 Please use specified bulb in case of damage or fire. 

Specified bulb

halogen 

12V

 /20W or 6V/30W

LED         

3.4V/3W LED(According to outfits

(1)  shut off the power

switch to "

" side

,unplug the power 

socket

 

(2)  waiting about 30 minutes until lamp box is cold. 

 Please be sure that the bulb is cool, then follow by the nest operations. 

(3)  Lay aside the microscope reliably, unscrew the knurled thumb screw of the lamp 

housing cover on the underside of base. 

(4)  Pull the knurled thumb screw gently outward, turn out lamp holder flap. 
(5)  Confirm the bulb is cold down, Pull out the old bulb carefully

 

(6)  Hold a new bulb with silk cloth to avoid fingerprint and dust affect bulb brightness and 

service life, and insert fully the contact pins into the bulb socket

 

  Before replace bulb

confirm if bulb connected strongly with the power cable in case of 

bad contact. 

  Plug fully contact in bulb socket, or will result in short circuit. 

(7)  Push back the lamp holder flap, tighten the knurled thumb screw. 

  After working 10 hours

suggesting the suspension of using about half an hour for 

continue using. 

B

Replacement of LED

 

      Please contact us. 

Содержание A15.2603

Страница 1: ...A15 2603 Polarizing Microscope Instruction Manual To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly be fore start to use the instrument...

Страница 2: ...CONTENT Attentions 2 1 Parts Name 3 2 Performance Parameter 4 3 Installation And Carry 5 4 Operation And Use 6 5 Maintenance 7 6 Troubleshooting 9 Appendix outfits 10...

Страница 3: ...e alcohol gasoline paper and other combustibles near the instrument to prevent a fire 5 Notes on replacing the bulb The instrument used with 12V 20W or 6V 30W halogen lamp of G4 LED model with 3 4V 3W...

Страница 4: ...g Knobs Brightness control knob 4 Compensator Built in Polarizer Clamps Analyzer driving lever Stand Bertrand Lens Nosepiece Center Adjustable rotating Stage Diopter Adjusting Ring Midbody Power Fuse...

Страница 5: ...ar Trinocular head Interpupillary Distance 48 mm 75 mm Dioper Adjusting 5 With antifungus device 3 Midbody Analyser 360 Division 2 30 With locking device Bertrand lens with center Adjustable Combining...

Страница 6: ...down nosepiece s dust cover take out objective and install in sequence for use 10 Rotate Coarse focusing knob lower rotating stage screwing 10Xobjective in objective hole which on the nosepiece s out...

Страница 7: ...ion 3 gypsum slip first class red Orthogonal polarizing appear first class purplish red interference colours 4 Mica 4 slip Orthogonal polarizing appear first class grey colours 5 Quartz wedge Orthogon...

Страница 8: ...ement of halogen Please use specified bulb in case of damage or fire Specified bulb halogen 12V 20W or 6V 30W LED 3 4V 3W LED According to outfits 1 shut off the power switch to side unplug the power...

Страница 9: ...in fuse tube socket again 5 5 Not in use 1 Please shut off the power covered by dust cover keep in dry and cool place 2 The eyepiece and objective should be placed in dry containers such as moisturep...

Страница 10: ...objective to dip in immersion oil Clean it Immersion oil isn t used for 100 objective oil Use immersion oil Immersion oil doesn t meet the requirement Use immersion oil supplied by us There is bubble...

Отзывы: