Operation And Use

 

Bright field operation process instruction 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

△ 

When inspecting to confirm polarizing direction

please finding a cleavage black mica 

from mineral    slice and place it in center of field ,then rotate stage again

Stop rotating 

when black mica’s color become very dark, the direction of cleavage seam just present 
the polarizing direction

 

△ 

In  positive  cross  polarized  light,  if  interference  colours  of  mineral  slice  is  first  class 
grey ,after insert gypsum (

λ

) in optical path, please change to second class blue-green if 

homology radius is parallel; if synonyms radius is parallel, change first class grey to first 
class orange color. 

△ 

In  positive  cross  polarized  light

if  interference  colours  of  mineral  slice  is  first  class 

purplish red ,after insert mica(

λ

/4) in optical path, please change to second class blue; if 

synonyms radius is parallel, lower to first class orange color. 

△ 

Analyzer  change-over  handle  can  convert  optical  path  ,  switching  observation  from 
mono polarized light to positive cross polarized light. 

△ 

Midbody

analyzer

compensator and Bertrand lens bearing. 

1. Compensator socket: can plug in gypsum

λ

Mica

λ

/4

or Quartz wedge. 

2. Bertrand lens conversion inserted  block: pulling  it can  convert  Bertrand  lens in  light             

path to make conoscopic observation. 

3. gypsum

λ

slip

first class red

Orthogonal polarizing appear first class purplish red     

interference colours. 

4. Mica

λ

/4

slip

Orthogonal polarizing appear first class grey colours. 

5. Quartz  wedge

Orthogonal  polarizing  appear  first  class  to  four  class  interference 

colours from thin to thick. 

Turn power switch

rotating screw to adjust 

brightness of Field.

 

Place specimen on stage, moving to the light path 
until to observe position

gently press clamps, 

moving to slice’s both end then fixed. 

Rotate nosepiece

put objective10

× 

in light path

rotate Coaxial Coarse And Fine Focusing handle 
wheel to focusing until legible image.(adjust 
tension ring to adjust coaxial coarse and fine 
focusing circle’s tension. 

Polarizing direction is reseted in horizontal level. 
(can through complete 

cleavage 

black mica 

sheet to define polarizing direction. 

When polarizer and analyzer located in dial 0°

is 

orthogonal each other and FOV will be dark(no 
sample),if not, should check polarizing position. 

According to mineral slice interference colors

plug 

in slider

 

1. Switch on the power, adjust brightness.

 

2. Place Mineral Specimen. 

3. Put 10X objective into optical path and 

focus. 

4. Define polarizing direction. 

5.Check polarizer and analyzer is 

orthogonal. 

6. Use compensator slider. 

Observing 

7. Observe with Bertrand lens   

Conoscopic  observation  should  use  high  power 
eyepiece

push  in  Bertrand  lens  when  in 

orthogonal  polarization  ,to  determine    mineral’s 
optical sign and optic axial angle

 

Содержание A15.2603

Страница 1: ...A15 2603 Polarizing Microscope Instruction Manual To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly be fore start to use the instrument...

Страница 2: ...CONTENT Attentions 2 1 Parts Name 3 2 Performance Parameter 4 3 Installation And Carry 5 4 Operation And Use 6 5 Maintenance 7 6 Troubleshooting 9 Appendix outfits 10...

Страница 3: ...e alcohol gasoline paper and other combustibles near the instrument to prevent a fire 5 Notes on replacing the bulb The instrument used with 12V 20W or 6V 30W halogen lamp of G4 LED model with 3 4V 3W...

Страница 4: ...g Knobs Brightness control knob 4 Compensator Built in Polarizer Clamps Analyzer driving lever Stand Bertrand Lens Nosepiece Center Adjustable rotating Stage Diopter Adjusting Ring Midbody Power Fuse...

Страница 5: ...ar Trinocular head Interpupillary Distance 48 mm 75 mm Dioper Adjusting 5 With antifungus device 3 Midbody Analyser 360 Division 2 30 With locking device Bertrand lens with center Adjustable Combining...

Страница 6: ...down nosepiece s dust cover take out objective and install in sequence for use 10 Rotate Coarse focusing knob lower rotating stage screwing 10Xobjective in objective hole which on the nosepiece s out...

Страница 7: ...ion 3 gypsum slip first class red Orthogonal polarizing appear first class purplish red interference colours 4 Mica 4 slip Orthogonal polarizing appear first class grey colours 5 Quartz wedge Orthogon...

Страница 8: ...ement of halogen Please use specified bulb in case of damage or fire Specified bulb halogen 12V 20W or 6V 30W LED 3 4V 3W LED According to outfits 1 shut off the power switch to side unplug the power...

Страница 9: ...in fuse tube socket again 5 5 Not in use 1 Please shut off the power covered by dust cover keep in dry and cool place 2 The eyepiece and objective should be placed in dry containers such as moisturep...

Страница 10: ...objective to dip in immersion oil Clean it Immersion oil isn t used for 100 objective oil Use immersion oil Immersion oil doesn t meet the requirement Use immersion oil supplied by us There is bubble...