aperture,  not  use  for  adjusting  brightness.  Usually,  for  most  specimen,  it  can  be  obtained  a 

good , enough contrast image if reducing the diaphragm opening to 70- 80% of the N.A. of 

the objective in use. 

·Adjusting the center of the aperture diaphragm (iris diaphragm): 

The center of the condenser's aperture diaphragm(iris diaphragm) has been pre-adjusted at 

the factory, and should not require readjustment. Push the Bertrand lens in or take away the 

ocular,  look  into  the  tube,  you  will  see  a  bright  circle.  Changing  the  diaphragm  opening 

slowly, you can observe the circle is changed coaxially. 

5

Orthogonal polarized light viewing

 

(1)When the polarizer is in the center of the optical path, its polarized direction is east-west at 

0°.The analyzer is located in the center of the optical path all the time, and marked with0°- 

90° graduation. Swing the analyzer lever, let the graduation center to the 90° scale line.    So 

the  two  polarized  prism  is  orthogonal  exactly.  That  means  the  polarized  direction  of  the 

polarizer is east-west, and the direction of the analyzer is south-north. 

(2)When the polarizer is get out of the optical path, the analyzer will still stay in the center of 

the optical path. The instrument is in the single polarizing observation. 

6

Convergent light observation

 

The  convergent  light  observation  is  often  use  the  high  magnification  objective.  Under  the 

orthogonal polarized light state, push in the bertrand lens, and rise up the condenser, then you 

can observe the convergent light performance of the specimen.   

7

Optical Compensator: 

          Constant  optical  compensator  just  like1/4λ,  λ  slip  and  quartz  wedge  variable  optical 

compensator is equipped. After the specimen focus, separately insert 1/4λ slip (micaceous), λ 

slip (gypsum) and quartz wedge into the slot under the analyzer at 45°, and you can observe 

first  class  gray  interferogram,  first  class  red  interferogram  and  the Ⅳ  class  avriable  optical 

interferogram of the specimen correspondingly. 

8

Adjusting the illumination system

 

The  illumination  has  been  pre-adjusted  at  the  factory.  If  need  to  change  the  lamp,  please 

follow  the  bellow  step:  First,  pull  out  the  button  under  the  base,  and  carry  down  the  bulb, 

replace  it  with  a  new  one.  Second,  loose  the  light  source  adjusting  bolt,  and  adjust  the 

filament  into  the  center  of  the  viewing  field,  then  screw  down  the bolt,  so  you have  a  best 

illumination environment for observation.. 

   

Ⅵ.MAINTENANCE   

1

 

Carefully  open  the  box,  avoid  the  accessories  dropping  to  ground  and  being  damaged,       

for example, the lens. 

2

 

All the lenses have been well checked and adjusted. it is forbidden to disassembly them 

by yourself. 

3

 

The nosepiece and coarse/fine focus unit have a compact and precise frame, please don’t 

disassembly them as possible as you can. 

4

 

Keep the instruction clean, wipe dust regularly. Pay attention to avoid contaminating the 

optical elements especially.   

5

  Place the instruction in a cool, dry position, when end use the microscope, remember to     

Содержание A15.1016

Страница 1: ...1 A15 1016 A15 1017 Polarizing Microscope Instruction Manual Please read this manual carefully before using the microscope and let it s along with the instrument...

Страница 2: ...r plate quartz wedge wedge wedge wedge Strain Free Objective Rotatable Stage Stage rotation clamping knob Polarizer Condenser Bertrand Lens Microscope Body Coarse Focus Knob Fine Focus Knob Power Swit...

Страница 3: ...32 dry 60 optional 0 85 0 17 2 969 0 185 dry 3 Crosshair Scale Ocular KIND MAGNIFICATI ON FOCUS MM LINEAR FIELD OF VIEW MM PLAN LENS 10 24 95 18 4 Sum Magnification OCULAR 10 10 10 10 OBJECTIV 4 10 40...

Страница 4: ...iece Quadruple Nosepiece Center Adjustable Optical Compensator slip first class red 1 4 slip quartz wedge blank slip Illumination Halogen Lamp 6V20W spare Lamp spare BGX 1 20 0 5A Fuse 2 2 Eyepiece Ey...

Страница 5: ...nto the holder stably and safely and return the top back to finish Don t touch the Halogen bulb with your hands Please do the replacement with your glove or the lamp protected cover If you leave finge...

Страница 6: ...l axis coincident with each other The nosepiece is center adjustable the objective optical axis and stage has been pre adjusted coincident at the factory and should not require readjustment If they ar...

Страница 7: ...ogonal polarized light state push in the bertrand lens and rise up the condenser then you can observe the convergent light performance of the specimen 7 Optical Compensator Constant optical compensato...

Страница 8: ...in pure alcohol or xylene Not use organic solvent to wipe the non optical elements when must clean please use the soft detergent note that the alcohol and the xylene are all burned easily do not let...

Отзывы: